Operation
Cleaning & Disinfecting
Inspection
NOTE
Keep up−to−date maintenance records using the form on page 1−34.
Parts, Service or Technical Assistance:
Contact Stryker Customer Service at 1−800−327−0770 or
Stryker Medical
3800 E. Centre Ave.
Portage, MI 49002
ATTN: Customer Service
CHECKLIST
All fasteners secure (reference all assembly drawings)
All welds intact, not cracked or broken
No bent or broken tubing or sheet metal
No debris in wheels
All wheels secure, rolling and swivelling properly
Optional wheel lock holds wheel securely when on and clears wheel when off
Siderails move and latch properly
Backrest operating properly
Optional accessories intact and operating properly
Height positioning latch functioning properly
Cot secure in each height position
Undercarriage folds properly
Breakaway head section operating properly
Safety bar operating properly
Foot rest operating properly
No rips or cracks in mattress cover
Body restraints intact and working properly
Lubricate base tubes (optional)
Serial No.______________
Completed By:_________________________________
Preventative Maintenance
Schedule
Each use.
For 1−25 calls per month, in-
spect cot every 6 months.
For 26−200 calls per month, in-
spect cot every 3 months.
For 201+ calls per month, in-
spect cot monthly.
1−31
Procedure
See page 1−30.
See below for checklist.
Date:_____________