Preventive Maintenance - Stryker Medical T/Pump TP700 Manuel D'utilisation

Système de thermothérapie localisée
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by qualified personnel.
WARNING
Risk of electric shock. Disconnect power before servicing or cleaning the T/Pump.
Note: Change the water monthly or more often depending on use.
Inspect the following items:
Line cord, the plug, and the housing for cuts or cracked insulation and quality of the strain relief, if damaged
replace
Membrane panel for cuts or cracks, if damaged replace
Plastic parts on the pump for cracks
Connectors for cracks and missing O-rings, replace connectors if necessary
Hose connections and connectors function properly
Pads and hoses are free from tears, cracks, and water leaks
Refer to the T/Pump Maintenance Manual Functional Checks section that and complete during preventive maintenance.
Inspection
Value
Measure current
leakage <100 µA
Product serial number:
Completed by:
Date:
www.stryker.com

Preventive maintenance

Okay
101515 REV E
Action Needed