Preventative Maintenance
BASE LUBRICATION
NOTE
The
MX−PRO has been designed to operate without the need for lubrication. Tri−Flowt
with TeflonR lubricant may be used to reduce the force required to raise the undercarriage and to minimize
wear. Do not use silicone, WD−40t, or lithium grease. They could harm the moving parts of the cot.
FOOT END
B
APPLY LUBRICATION
A
Figure 37 − Base Lubrication Locations
1. Place the cot in the highest position.
2. Turn the cot upside down with the base legs facing up.
3. Using the plastic applicator tube, spray Tri−Flowt with TeflonR lubricant (Stryker part number
6082−199−12) in the gap between the upper (item A) and lower lift tubes (item B) on both foot end legs.
4. Let the cot sit for 5 minutes.
5. Turn the cot back over. Raise and lower the cot a few times to work the lubricant through the base tubes.
6. Wipe any excess lubricant from the base tubes.