English
Press the
to adjust the numerical display to the desired value ;
Press the
The instrument then returns to the conductivity measurement mode.
NOTE: adjustment range 0 to 8.
6. TABLE: CONDUCTIVITY OF POTASSIUM
6.1 CONDUCTIVITY VS. TEMPERATURE
TEMPERATURE
°C
15
16
17
18
19
20
21
22
23
24
25
66
or
key to validate.
CHLORIDE SOLUTIONS
Dilution to 1/10
Dilution to 1/50
-1
KCL 0.1 mol.L
KCL 0.02 mol.L
Conductivity
mS/cm
10.410
10.670
10.930
11.190
11.430
11.700
11.960
12.220
12.470
12.730
12.970
key
Dilution to 1/100
-1
KCL 0.01 mol.L
Conductivity
mS/cm
2.242
2.293
2.347
2.398
2.451
2.500
2.551
2.604
2.659
2.710
2.769
MPC 350 Multiparameter meter
Dilution to 1/1000
-1
KCL 0.001 mol.L
Conductivity
µS/cm
1,147
1,174
1,199
1,224
1,250
1,279
1,305
1,331
1,359
1,387
1,412
-1
Conductivity
µS/cm
119.1
121.9
124.5
127.1
129.6
132.5
135.3
138.1
140.9
143.8
146.5