The numerical display indicates the cell coefficient.
(The coefficient is expressed in cm: ratio of the surface area of the two plates of
the cell to the distance between them).
The instrument then returns to the measure.
5.1.3
VIEWING THE CELL COEFFICIENT:
Press the
Then:
The numerical display indicates
the cell coefficient:
5.1.4
ADJUSTING THE TEMPERATURE COEFFICIENT:
Because the solutions evolve differently as a function of temperature, it is
necessary to set the temperature coefficient (expressed in % per degree) so as to
refer the conductivity to 25°C whatever the actual measurement temperature.
Press the
The alphanumeric display indicates:
Then:
The numerical display indicates the current value of this parameter.
(The factory setting is 2.2%/°C, the commonest value).
MPC 350 Multiparameter meter
key ;
key.
1.01 COEF CEL
COEF CEL
6-7 et 8
English