7. IF SOMETHING GOES WRONG

PROBLEM
No display
No external temperature reading
Conductivity fails to stabilize in all
solutions measured
No variation of conductivity
Displays "1999" and "SOS * S/cm"
in conductivity
Affiche « 1999 » en conductivité
The pH fails to stabilize with all
solutions, or the display indicates:
"14.00" and "* SOS * pH*"
In pH, a value close to pH 7.00 is
displayed in all solutions.
In mV, displays:
"1999" and "* SOS * mV *"
74
Check the 230V supply.
Check the connection of the power supply unit to
the power connector.
Check the connection of the temperature probe
Check the value of the probe
(107Ω at 20°C)
Check that the conductivity cell is correctly
immersed.
Check the connection of the cell.
Check the condition of the cell.
If necessary, replace it, or perform tests with pure
resistances.
For 1kΩ the value displayed 1000µS.
You are over the range limit.
Switch to the next higher range.
Check the immersion of the electrode.
(Reference jonction, glass bulb)
Check its connection.
Check the condition of the electrode.
(Glass bulb)
Check the immersion of the electrode.
Check its connection.
MPC 350 Multiparameter meter
ADVICE