Press the
If the calibration is correct, the instrument automatically returns to the pH
measurement mode.
Note: calibration with a single buffer does not adjust the slope of the
electrode, but only the asymmetry potential (asymmetry accepted between
pH6 and pH8)
3.3.2.3
Manual calibration with 1 or 2 buffers:
Example: buffers 7 and 9 with manual temperature correction.
Check the temperature of the buffer solutions and if necessary adjust the manual
temperature (§ 3.2.3 Manual correction)
Electrode rinsed, placed in buffer no. 1 (pH 7.00):
Press the
The alphanumeric display alternates between:
Press the
The alphanumeric display alternates between:
Press the
56
key.
key.
key.
key for manual calibration with 1 buffer
AUTO 1
MAN 2
1 ou 2 T
MPC 350 Multiparameter meter