6.3.2
ADJUSTMENT OF THE RANGE LIMITS (0/100 AND 100/100) OF THE
RECORDER OUTPUT:
Press the key corresponding to the parameter of which you want to modify the
limits.
Example: pH
Press the
The MPC 350 changes to the pH measurement mode.
Press the
The alphanumeric display indicates successively:
0/100 corresponds to 0V :
The numerical display indicates the present setting.
Press the
to adjust the numerical display to the desired 0/100 value.
Then press on the
The alphanumeric display indicates:
100/100 corresponds to 5V :
MPC 350 Multiparameter meter
key.
key.
or
key to validate.
*
ENR pH
0/100
key
100/100
English
69