TEMPERATURE MEASUREMENT

Press the
The display unit indicates the temperature:
The alphanumeric display indicates:
or, in the manual temperature mode:
3.2.2

AUTOMATIC CORRECTION

Place the temperature probe in the liquid to be tested. Automatic correction
replaces the action of the manual corrector, which is then inoperative.
It must not be forgotten that the response of the probe is not instantaneous, and
that the response time depends on the temperature gradient.
After stabilization, the temperature reading can be slightly different from that given
by a reference thermometer.
If this happens, the Pt 100 probe can be calibrated as follows:
Example: displays 22.5°C instead of 20°C
Press the
The instrument changes to temperature calibration mode
50
key to switch to temperature measurement mode.
key
22.1
** TC **
*TC MAN*
22.5 ** TC **
22.5 CAL TEMP
6 – 7 and 8
MPC 350 Multiparameter meter