7. CHANGING THE BATTERY
The clamp must be completely disconnected before changing the battery and it
I
must not enclose any conductor.
Check that the switch is in the OFF position.
I
Undo the tool-release screw at the back of the clamp and slide the cover into the
I
handle extension.
Connect up the battery (9 V alkaline battery), then replace the cover.
I
8.1 SETTING ZERO
It is only necessary to set zero for measurements of DC current or for measurements
on AC with a DC component.
Connect the E6N clamp to a
multimeter (black lead to the
common and red lead to the
«Volt» input).
Place the function/range
switch of the multimeter on
the mV DC position. Choose
a measurement extent that
has a resolution of 10 mV.
Place the function switch
(located on the handle) on
position 1 A/V.
Using the thumbwheel, display on the multimeter a value as close as possible to 0.
8. OPERATING MODE
Black
20
Red
mV DC
25
9