5

Post-use Maintenance

Follow local rules, regulations, and guidelines regarding the reprocessing of devices.
Perform the following cleaning and disinfection quickly after treatment (within 1 hour) to remove any residue.
Do not perform steam sterilization the control unit, motor cord and AC adapter, AC power cord.
Do not lubricate the motor. It may cause overheating and product failure.
Do not use the following fluids to wipe, immerse or clean the product; strong/super acid water, strong acid/
alkaline chemicals, chlorine containing solutions, solvents such as benzine or thinner. It may cause color change
of resin part, or corrosion of the metal part of the NLZ Motor System.
Steam sterilization is recommended for the product. The validity of other sterilization methods (such as plasma
sterilization or EOG sterilization) is not confirmed.
For details on maintenance of the handpiece, check the Operation Manual of the handpiece.
Do not immerse the product in disinfectant or perform cleaning with an ultrasonic washer. Doing so may cause
product failure.
After each patient, maintain the product as follows.
5-1
Cleaning at point-of use (Motor)
Always wear protective gloves, a mask, and
1
protective goggles for safety purposes and to
minimize the risk of infection.
3
Remove the bur/file.
CAUTION
Turn off power by pressing the Power Key
2
38
2