Operation Procedure
The motor operation will be controlled by the foot pedal of the dental unit (also called Rheostat). Pressing the
9
pedal all the way down will run the motor up to the Max speed set at step 4
Memory setting
10
Press and hold the Memory Key for more than 2 seconds. After 2 short beeps the memory sign will stop blinking
and gear ratio and speed will be memorized in the selected program number (M1 through M6 is shown, so that
up to 6 programs can be stored in memory).
A blinking
indicates that the program is not stored in the memory yet.
4-2
Rotary Endo Mode (NLZ E only)
2
Select the program # - 6
programs can be selected
3
Select the gear ratio
according to the handpiece
1:1, 4:1, 6:1
8
Set the light intensity
Set the light intensity
Strong Moderate Weak
4
Select the proper speed recommended by
file manufacturers
Gear Ratio
Rotation Speed (min
1:1
100 – 6,000
4:1
30 – 1,500
6:1
20 – 1,000
OFF
-1
)
Setting range (min
-1
)
20 – 500:10 Increment
500 – 1,000:50 Increment
1,000 – 6,000:100 Increment
NOTICE
5
Select the proper torque recommended
by file manufacturers
Gear Ratio Torque Setting Range (N・cm)
1:1
4:1
6:1
24
6
Select the Auto-Reverse
Mode: AUTO STOP /
AUTO REVERSE STOP /
AUTO REVERSE
FORWARD
1
Select the Rotary Endo
Mode
7
Select the rotation
direction, FWD / REV
Setting range (N・cm)
0.3 – 3.0
0.1 Increment
0.8 – 6.0
0.4 Increment
1.0 – 1.2:0.2 Increment
1.0 – 6.0
1.2 – 6.0:0.6 Increment