EMC Information
Guidance and manufacturer's declaration - Electromagnetic Emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should
assure that it is used in such an environment.
Emissions test
RF emissions
CISPR11/EN55011
RF emissions
CISPR11/EN55011
Harmonic emissions
EN/IEC61000-3-2
Voltage fluctuations/flicker
emissions
EN/IEC61000-3-3
Guidance and manufacturer's declaration - Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should
assure that it is used in such an environment.
Immunity test
Electrostatic discharge(ESD)
EN/IEC61000-4-2
Electrical fast transient/burst
EN/IEC61000-4-4
Surge
EN/IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN/IEC61000-4-11
Power frequency
(50/60Hz) magnetic field
EN/IEC61000-4-8
NOTE: 'Ut' is the AC mains voltage prior to application of the test level.
(Electromagnetic Compatibility Information)
Compliance
Group 1
Class B
No applicable
No applicable
IEC/EN60601 test level
±(2,4)6kV contact
±(2,4)8kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut
(>95% dip in Ut)
for 0.5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 0.5 cycles
<5% Ut
(<95% dip in Ut)
for 5 sec
3 A/m
Electromagnetic environment – guidance
The product uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The product is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Compliance level
±(2,4)6kV contact
±(2,4)8kV air
±2kV for power supply lines
±1kV for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut
(>95% dip in Ut)
for 0.5 cycles
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 0.5 cycles
<5% Ut
(<95% dip in Ut)
for 5 sec
3 A/m
52
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the product
requires continued operation during power
mains interruptions, it is recommended
that the product be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.