Overheat Prevention
Following functions detect a risk of overheating from motor current and rotation speed and then slows
down and stops motor rotation by activating the protection circuit to prevent incidents or failure due to the
overheat of the handpiece or the motor.
<Motor Overheat Prevention Function>
This function stops the rotation of the motor when detecting
overload or a risk of abnormal heating of the motor.
"Warning" and "HE0" will be displayed along with alarm
sound and rotation of the motor will be slowed down.
Continued use will result in motor stop and Error Code "E8" will be displayed. (Refer to "7-1 Error Code")
Gear ratio
Rotation speed (min
1
1:5
2
1:1
3
1:1
4
1:1
5
1:1
6
4:1
1
1:1
2
1:1
3
1:1
4
1:1
5
1:1
6
1:1
-1
)
Torque (N・cm)
200,000
40,000
30,000
20,000
10,000
3,000
250
1.0
250
1.5
250
2.0
250
3.0
300
2.0
1,000
3.0
NOTICE
36
36
Auto reverse mode
AUTO REVERSE FORWARD
AUTO REVERSE FORWARD
AUTO REVERSE FORWARD
AUTO REVERSE FORWARD
AUTO REVERSE FORWARD
AUTO STOP