EMC Information
(Electromagnetic Compatibility Information)
Cables and accessories Maximum length
Link cable
AC adaptor cord
(Secondary side)
AC adaptor cord
(Primary side)
Motor cord
Recommended separation distances between portable and mobile RF communications equipment and the product
The product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
the user of the product can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the product as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output power
of transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance "d" in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where "P" is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Complies with
1.0m
RF emissions, CISPR11, EN55011
(Unshielded)
Electrostatic discharge (ESD):
Electrical fast transient/burst:
5.0m
Surge:
(Unshielded)
Voltage dips, short interruptions and voltage variations on power supply input
2.0m
lines:
(Unshielded)
Power frequency(50/60Hz) magnetic field: EN/IEC61000-4-8
2.2m
Conducted RF:
(Unshielded)
Radiated RF: EN/IEC61000-4-3
Separation distance according to frequency of transmitter m
150MHz to 80MHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
EN/IEC61000-4-5
EN/IEC61000-4-11
EN/IEC61000-4-6
80MHz to 800MHz
d=1.2 √P
0.12
0.38
1.2
3.8
12
54
Class B / Group 1
EN/IEC61000-4-2
EN/IEC61000-4-4
800MHz to 2.5GHz
d=2.3 √P
0.23
0.73
2.3
7.3
23