3. Remove fuse socket caps (#1 and #2, Figure 5-4-1 or 5-4-2) with a screwdriver. The cap
is not a screw; therefore, do not try to unscrew it or the cap may be damaged. Just push in
the cap and turn slightly to unlock it and use hand to pull it out.
Figure 6-2
4. Replace blown fuse(s) with new one(s) according to the specification below.
NUVO Item ID
EF0002
EF0003
5. Push and turn slightly to lock the fuse socket caps.
6.9
System Reset
Preliminary check:
1. Unplug AC and the AC LED that is on the base panel should be off, and vice versa.
2. While the AC is plugged in, turn on hand pendant, and its panel should be lit.
3. The battery power indicator should come on after the hand pendant is turned on.
4. While the hand pendant is on, perform the following.
For B2AV1000S(K)(P):
1. Flip the override control enable toggle switch to the "on" (up) position. It is the first toggle
switch from the right.
2. Wait 5 seconds, and the hand pendant should be turned off, which means the system has
been reset.
Specifications
6*32 3A/250V (time-lag)
6*32 5A/250V (time-lag)
6-9
B2AV1000 Series Surgical Table
IM-B2AV1000-E01
AC Voltage
220V~240V
110V~120V