Level Button Operation; Normal Orientation Operation; Reverse Orientation Operation; Battery Level Indicator - Nuvo B2AV1000-1 Manuel

Table des Matières

Publicité

4.
When the table begins to move on its own while using the hand pendant, press the OFF
button right away and call for service.

5.2.1.1 Level Button Operation

Press Level button (#27, Figure 5-2) until tabletop has reached level and then let go of
the button.

5.2.1.2 Normal Orientation Operation

Press Normal button (#4, Figure 5-2) and its green LED will be lit. All buttons will
automatically switch to normal orientation.

5.2.1.3 Reverse Orientation Operation

Press Reverse button (#6, Figure 5-2) and its green LED will be lit. All buttons will
automatically switch to reverse orientation.

5.2.1.4 Battery Level Indicator

Besides indicating level of battery power, it also indicates the need to change the
battery or batteries when all three LEDs within the battery symbol are flashing.

5.2.1.5 Interlock LED

As a protective feature, when the Interlock indicator (#17, Figure 5-2) is lit (buzzer
beeping at the same time), it means one of movements (table down, Trendelenburg,
Reverse Trendelenburg, tilts, leg down, back down or slide out) has been stopped for
running into something and the button of the above said movements is disabled. It
could also mean that one of the above said movements has reached its maximum
distance or angle.

5.2.1.6 Service LED

While operating Hand Pendant, if Service indicator (#18, Figure 5-2) is lit (buzzer
beeping at the same time), it could mean override system is on. Turn it off and the
Service LED should disappear. Other than that, when Service LED is on while using
the hand pendant, then a problem has occurred that renders certain component(s)
unusable. At this time, the 10-segment LED Battery Indicator (#8, Figure 5-4-1 or #8,
Figure 5-4-2) could display Service Code. For other events, which the Service LED is
lit only while a button is pressed; press the button while looking for the Service Code
display. Let go of the button to exit from the Service Code display. Inform your
finding to the NUVO Sales or Service Representatives. (Refer to Table 5-1 for
description of Service Codes.)
5-10
B2AV1000 Series Surgical Table
IM-B2AV1000-E01

Publicité

Table des Matières
loading

Table des Matières