Display/Code
No signal in more than one angle
10
sensor
11 Not Used
Motor is running as soon as the hand
12
pendant is turned on.
13 When rotary valve keeps turning but
receives no signal from U1~U4 of
ASM0027 (B2AV1000(K))
Rotary valve position number not
14
matching position code
(B2AV1000(K))
Rotary valve position code correct but
15
stops too early, and fails to position
after programmed compensation.
(B2AV1000(K))
Rotary valve position code right but
16
stops too, and fails to position after
program compensation.
(B2AV1000(K))
Tabletop rotates laterally or
17
longitudinally without their respective
button being pressed.
(B2AV1000S(K)(P))
Not Used
18
19 Not Used
Not Used
20
Main batteries is not being charged
21
(i.t.)*
Not Used
22
Symptom
5-12
B2AV1000 Series Surgical Table
IM-B2AV1000-E01
Cause
Loosed or damaged ASM0028
connection (P17)
Main PCB or Driver PCB is
damaged.
1. Short circuit within the angle
system.
2. Rotary valve motor damaged
3. ASM0023 driver PCB
damaged
4. ASM0027 or its cable
damaged
ASM0027 or its cable damaged
Timing parameter is set too low
Timing parameter is set too high
Faulty ASM0022 driver PCB
1. Loose battery connector
2. Battery reached life span or
over discharged
3. Broken PCB (ASM0021)