1. Push Hand Pendant to firmly clip it on
an appropriate side rail. Do not only
hang it on the side rail.
Note:
When removing the Hand Pendant
from the side rail, pull it straight out.
Do not twist from side to side, or the
metal clip could be damaged.
Remarque:
Lorsque vous retirez le boîtier de commande manuelle du rail latéral, tirez-le vers vous. Ne
le tournez pas de gauche à droite, car cela pourrait endommager l'attache en métal.
2. Locate the receptacle on column top
under the seat section.
3. Insert the connector all the way into the
receptacle, and then tighten its screw
to prevent the connection from
becoming loose.
4.3 Lock Table in Place
SHOCK HAZARD: Monitor the electrically conductive properties of floor locks.
Measure their resistance level according to the requirements of NFPA 99 before placing
table in use. Subsequent routine testing should be performed at least once every three
months.
RISQUE DE DÉCHARGE ÉLECTRIQUE: Surveillez les propriétés de conductivité
électrique des dispositifs de verrouillage au sol. Mesurez leur niveau de résistance
selon les exigences de la norme NFPA 99 avant de commencer à utiliser la table. Des
tests de routine doivent être effectués au moins une fois tous les trois mois.