B2AV800 Series Surgical Table
IM-B2AV800-E01
B2AV800 Series
SURGICAL TABLE
INSTRUCTION MANUAL
Date of Issue: July 26, 2012
This manual is intended for use by NUVO customers and distributors only. All rights reserved.

Publicité

Table des Matières
loading

Sommaire des Matières pour Nuvo B2AV800-1

  • Page 1 B2AV800 Series Surgical Table IM-B2AV800-E01 B2AV800 Series SURGICAL TABLE INSTRUCTION MANUAL Date of Issue: July 26, 2012 This manual is intended for use by NUVO customers and distributors only. All rights reserved.
  • Page 3 B2AV800 Series Surgical Table IM-B2AV800-E01 TABLE OF CONTENTS Section Paragraph Page REVISION HISTORY REVISION DATE OF ISSUE REASON FOR CHANGE July 01, 2009 New Release October 3, 2011 Change of Electrical Circuitry LIST OF AFFECTED PAGES DIRECTION CHAPTER PAGE REVISION IM-B2AV800-E IM-B2AV800-E 6, 8...
  • Page 4 B2AV800 Series Surgical Table IM-B2AV800-E01 TABLE OF CONTENTS Section Paragraph Page Dear Technician or User, Before installing and using this table, please read these installing and operating instructions thoroughly. This manual also describes the maintenance procedures. Technicians installing and/or servicing this table should be familiar with the warnings, cautions, and instructions contained herein.
  • Page 5: Table Des Matières

    B2AV800 Series Surgical Table IM-B2AV800-E01 TABLE OF CONTENTS Section Paragraph Page SUMMARY OF WARNINGS AND CAUTIONS………………………1-1 SYMBOLS, TOOLS AND MATERIALS ........... 2-1 GENERAL INFORMATION .............. 3-1 Product Description .............. 3-1 Product Classification and Identification ....... 3-2 Requirements ............... 3-2 3.3.1 Operating Environment Requirements .........
  • Page 6 B2AV800 Series Surgical Table IM-B2AV800-E01 TABLE OF CONTENTS Section Paragraph Page OPERATING INSTRUCTIONS ............5-1 Pinch Point Warnings ............5-7 Table Positioning ..............5-9 5.2.1 Hand Pendant Operation ............. 5-9 5.2.1.1 Level Button Operation ............5-10 5.2.1.2 Normal Orientation Operation ..........5-10 5.2.1.3 Reverse Orientation Operation ..........
  • Page 7 B2AV800 Series Surgical Table IM-B2AV800-E01 TABLE OF CONTENTS Section Paragraph Page Quarterly Maintenance ............6-5 Lubrication Schedule ............6-6 Schedule for Other Maintenance .......... 6-6 Battery Charging Procedures ..........6-7 Fuse Replacement ............. 6-8 System Reset ..............6-10 SPARE PARTS ................7-1 CIRCUIT DIAGRAMS ...............
  • Page 8 B2AV800 Series Surgical Table IM-B2AV800-E01 TABLE OF CONTENTS Section Paragraph Page This page intentionally left blank...
  • Page 9: Summary Of Warnings And Cautions

    Please note that these precautions are not comprehensive. It is not possible for NUVO to know, evaluate and advise maintenance departments about all of the necessary maintenance procedures or the potential dangers of each procedure.
  • Page 10 (if applicable). Do not use worn or damaged accessories. PERSONAL INJURY HAZARD: Only recommended cleaners/disinfectants should be used on NUVO surgical tables. Do not use alcohol, which does not have sufficient cleaning/disinfecting properties, or phenolics, which when inadequately rinsed off could...
  • Page 11 B2AV800 Series Surgical Table IM-B2AV800-E01 WARNING: burn the patient’s skin. PERSONAL INJURY HAZARD: If Clark socket is not securely attached to the side rail, it could fall on and cause injury to a person’s foot. INTERFERENCE HAZARD: Do not place this surgical table near other equipment that will generate electromagnetic interference or other types of interference.
  • Page 12 B2AV800 Series Surgical Table IM-B2AV800-E01 AVERTISSEMENT: catalogue de la série B2AV800, veuillez contacter votre représentant local NUVO. RISQUE DE BASCULEMENT: La limite nominale de poids des patients est testée conformément à la norme IEC 60601-2-46, sous-clause 21.101. Le modèle de calcul standard de répartition du poids est basé...
  • Page 13 AVERTISSEMENT: être autorisé à retirer les couvercles. RISQUE D'INSTABILITÉ: L'utilisation de la table NUVO et de ses accessoires ou d'accessoires fabriqués et vendus par d'autres compagnies pour d'autres fins que celles indiquées peut entraîner des blessures au patient ou à l'opérateur, ainsi que des dommages à...
  • Page 14 CAUTION:  Only recommended cleaners/disinfectants should be used on NUVO tables. Do not use alcohol, which does not have sufficient cleaning/disinfecting properties, or phenolics, because when it is inadequately rinsed off, it could burn patient’s skin. ...
  • Page 15 MISE EN GARDE :  Seuls les nettoyants et désinfectants recommandés doivent être utilisés sur les tables NUVO. N'utilisez pas d'alcool pour procéder au nettoyage de la table, car ses propriétés de nettoyage et de désinfection sont insuffisantes et n'utilisez pas de composés phénoliques qui risquent de brûler la peau du patient s'ils sont mal rincés.
  • Page 16  Avant d'utiliser de l'équipement chirurgical de haute fréquence, tel que des défibrillateurs cardiaques ou des moniteurs à défibrillateur cardiaque, veuillez vous référer à leurs manuels d'utilisation.  Utilisez uniquement un couvre-matelas NUVO afin d'éviter les risques de décharges électrostatiques.  Si le cordon d'alimentation est endommagé, remplacez-le immédiatement ou utilisez la source d'alimentation électrique interne de la table jusqu'à...
  • Page 17 B2AV800 Series Surgical Table IM-B2AV800-E01 MISE EN GARDE : tordre, sinon les broches internes du réceptacle risquent de se plier et d'être inutilisables.  Puisqu'un choc sur les couvre-joints de la table risque de les endommager, gardez les brancards et autres objets lourds à l'écart des couvre-joints. ...
  • Page 18 B2AV800 Series Surgical Table IM-B2AV800-E01 This page intentionally left blank 1-10...
  • Page 19: Symbols, Tools And Materials

    B2AV800 Series Surgical Table IM-B2AV800-E01 Section 2 SYMBOLS, TOOLS AND MATERIALS SYMBOLES, OUTILS ET MATÉRIELS Table 2-1: Following symbols may be seen on your table or controls. Tableau 2-1: Vous pouvez voir les symboles suivants sur la table ou sur les systèmes de commande.
  • Page 20 B2AV800 Series Surgical Table IM-B2AV800-E01 Definition Symbols (Symboles) (Définition) Powered by AC when LED is on, by DC when off (Alimenté par le courant CA lorsque le voyant DEL est allumé, et par la batterie interne lorsqu'il est éteint) Full Battery (Batterie chargée) When all three LEDs are flashing, batteries should be replaced.
  • Page 21 B2AV800 Series Surgical Table IM-B2AV800-E01 Definition Symbols (Symboles) (Définition) On Hand Pendant, it turns the Hand Pendant into Sleep Mode. On Override panel, it disables the override system. (Sur le boîtier de commande manuelle, ce bouton active le boîtier de commande manuelle.
  • Page 22 B2AV800 Series Surgical Table IM-B2AV800-E01 Definition Symbols (Symboles) (Définition) Tilt Left (Incliner à gauche) Tilt Right (Incliner à droite) Back Down (Baisser le dossier) Back Down (Baisser le dossier) Back Up (Elever le dossier) Back Up (Elever le dossier) Leg Down (Baisser les jambes) Leg Down (Baisser les jambes)
  • Page 23 B2AV800 Series Surgical Table IM-B2AV800-E01 Definition Symbols (Symboles) (Définition) Slide Toward Leg (Glisser vers les jambes) Slide toward Leg (Glisser vers les jambes) Reflex (Flexion inverse) Flex (Flexion) Zero Line Shift (Décalage de la ligne zéro) Table 2-2: Tools and Materials Following tools and materials may be needed during installation and/or maintenance.
  • Page 24 B2AV800 Series Surgical Table IM-B2AV800-E01 Item Name and Information Picture T005 Phillips (crosshead) screwdriver (+) T006 Slotted (flathead) screwdriver (─) Grease: T018 MOLYKOTE 33 Medium...
  • Page 25: General Information

    There are six models in the B2AV800 Series as shown in the following table: Table 3-1: Model-Function Comparison Function DIP Switch (ASM0021) Tabletop Kidney Floor Model Sliding Elevator Locks No. 3 No. 4 B2AV800-1 (-2) manual ○ B2AV800K-1 (-2) manual ○ B2AV800S-1 (-2) manual ○ ○ B2AV800SK-1 (-2) manual B2AV800L-1 (-2) auto ○...
  • Page 26: Product Classification And Identification

    1. Plug the power cord into an equivalent electrical outlet marked “Hospital only” or “Hospital Grade” for effective electrical grounding. 2. B2AV800 Series surgical tables are shipped with standard electrical specifications and configurations. If a change needs to be made to the electrical system, contact NUVO for instructions on the correct procedure. Remarque: 1.
  • Page 27: Product Transportation And Storage

    2. Les tables chirurgicales de séries B2AV800 sont livrées avec des configurations et des spécifications électriques standards. Si des modifications doivent être apportées au système électrique, veuillez contacter NUVO pour les directives de procedure. Product Transportation and Storage The B2AV800 Series surgical table is heavy. Use the original package to transport and store the table.
  • Page 28: Technical Specifications

    B2AV800 Series Surgical Table IM-B2AV800-E01 Technical Specifications 3.6.1 Dimensions Figure 3-3: Dimensions 3.6.2 Weight (without package and accessories) Model B2AV800 B2AV800 B2AV800 B2AV800 B2AV800 B2AV800 Weight 396 (180) 407 (185) 416 (189) 427 (194) 431 (196) 442 (201) lb (kg) 3.6.3 Battery Capacity Battery Set: 24V, 14 Ah (Battery Specification: 12V/14Ah*2, per table)
  • Page 29: Imaging Coverage

    B2AV800 Series Surgical Table IM-B2AV800-E01 3.6.4 Imaging Coverage Figure 3-4: Models B2AV800, B2AV800K Normal Orientation Reverse Orientation Note: The imaging coverage is 14 inches wide. Remarque: La couverture d'imagerie est de 35.5 cm (14 po) de large Figure 3-4-2: Models B2AV800S, B2AV800SK, B2AV800LS, B2AV800LSK Normal orientation with table slid to head end Normal orientation with table slid to leg end...
  • Page 30: Load Capacity

    B2AV800 Series Surgical Table IM-B2AV800-E01 Reverse orientation with table slid to head end Reverse orientation with table slid to leg end Note: The imaging coverage is 14 inches wide. Remarque: La couverture d'imagerie est de 35.5 cm (14 po) de large 3.6.5 Load Capacity The nominal patient weight capacity is tested according to the Standard IEC60601-2-46...
  • Page 31: Duty Cycle

    Magnetic immunity regulation: 3A/m  Voltage Dips regulation: 30%/10mS, 60%/100mS, at 230VAC Accessories NUVO offers a wide range of accessories for its B2AV800 Series surgical tables. Please contact an NUVO sales representative for more information about B2AV800 Series surgical table accessories.
  • Page 32 B2AV800 Series Surgical Table IM-B2AV800-E01 This page intentionally left blank...
  • Page 33: Installation Instructions

    B2AV800 Series Surgical Table IM-B2AV800-E01 Section 4 INSTALLATION INSTRUCTIONS WARNING! AVERTISSEMENT! SHOCK HAZARD: Monitor the electrically conductive properties of the floor locks (all models) and casters (B2AV800S(K) and B2AV800(K)). Measure their resistance level according to the requirements of NFPA 99 before placing table in use. Subsequent routine testing should be performed at least once every three months.
  • Page 34: Install Ac Power Cord

    B2AV800 Series Surgical Table IM-B2AV800-E01 4.1 Install AC Power Cord 1. Move table to a desired location. 2. Plug power cord into socket (#4 of Figure 5-3), and secure it with Wire Collect Tie (#3 of Figure 5-3). 3. Route the cord in ways so that it could not be tripped over, and plug the cord’s 3-pronged hospital grade plug into an electrical outlet.
  • Page 35: Lock Table In Place

    B2AV800 Series Surgical Table IM-B2AV800-E01 1. Push Hand Pendant to firmly clip it on an appropriate side rail. Do not only hang it on the side rail. Note: When removing the Hand Pendant from the side rail, pull it straight out. Do not twist it from side to side, or the metal clip could be damaged.
  • Page 36: B2Av800Lsk And B2Av800Ls

    Base→ Patins de verrouillage au sol→Plancher. Par conséquent, les propriétés antistatiques de la table dépendent de l'utilisation des couvre-matelas recommandés par NUVO. 4.3 .1 B2AV800LSK and B2AV800LS 1 Press the Lock button (#2, Figure 5-2) of Hand Pendant and wait for the green LED to light up, the floor locks will become locked.
  • Page 37: Install Head Section

    B2AV800 Series Surgical Table IM-B2AV800-E01 4.4 Install Head Section 1. Loosen thumbscrews under the back section before attaching or removing the head section. 2. Insert connection rods of the head section into the back section. 3. Tighten the two thumbscrews to secure the head section.
  • Page 38: Performance Checklist

    B2AV800 Series Surgical Table IM-B2AV800-E01 4.6.1 Performance Checklist After completing installation procedures (4.1 to 4.5), installation technician shall verify the performance of table (also refer to 5.2.1 “Hand Pendant Operation”) according to the specifications and record the results of verification in the chart below: Test Item Specification Result...
  • Page 39: Voltage Setting Instructions

    B2AV800 Series Surgical Table IM-B2AV800-E01 4.7 Voltage Setting Instructions WARNING! AVERTISSEMENT! TABLE FAILURE HAZARD: This manual describes routine scheduled maintenance and services which will help keep this equipment safe and reliable. Only experienced personnel who are fully acquainted with this equipment should perform repairs and adjustments.
  • Page 40 B2AV800 Series Surgical Table IM-B2AV800-E01 : Tools and Materials Required Tools: T003 and T006 (Refer to Table 2-1 Procedures: 1. Remove protective sleeves from the column and lift up the base cover. 2. Unplug one of the connectors from the batteries. 3.
  • Page 41 B2AV800 Series Surgical Table IM-B2AV800-E01 Figure 4-6: Wiring of Terminals and Bridge Rectifier Main Control PCB (ASM0021) Bridge Rectifier Terminals 110 110 26V 24V 22V 20V 26V 24V 22V 20V...
  • Page 42 B2AV800 Series Surgical Table IM-B2AV800-E01 This page intentionally left blank 4-10...
  • Page 43: Operating Instructions

    3.6.5 “Load Capacity”). Do not exceed the limit. For maximum safe patient weight for any position other than those shown in the B2AV800 catalog, please contact your local NUVO representative. TIPPING HAZARD: The nominal patient weight capacity is tested according to the Standard IEC60601-2-46 Additional sub clause: 21.101.
  • Page 44 EXPLOSION HAZARD: Do not use this table with flammable anesthetics. INSTABILITY HAZARD: Using an NUVO surgical table and/or NUVO accessories or accessories manufactured and sold by other companies for other than the stated purpose could result in possible patient or user injury, and/or table, accessory, or other property damage.
  • Page 45 B2AV800, veuillez contacter votre représentant local NUVO. RISQUE DE BASCULEMENT: La limite nominale de poids des patients est testée conformément à...
  • Page 46 (électrique) de manière à éviter les risques de trébuchement. RISQUE D'EXPLOSION: N'utilisez pas la table avec des anesthésiques inflammables. RISQUE D'INSTABILITÉ: L'utilisation de la table NUVO et de ses accessoires ou d'accessoires fabriqués et vendus par d'autres compagnies pour d'autres fins que celles indiquées peut entraîner des blessures au patient ou à...
  • Page 47 Before using high-frequency surgical equipments, cardiac defibrillators or cardiac defibrillator-monitors, please refer to their instruction manual.  Use only NUVO’s mattress pads in order to avoid risk of electrostatic shock.  When the power cord is damaged in any way, replace it immediately or else use the table’s internal electrical power source until it can be replaced.
  • Page 48  Avant d'utiliser de l'équipement chirurgical de haute fréquence, tel que des défibrillateurs cardiaques ou des moniteurs à défibrillateur cardiaque, veuillez vous référer à leurs manuels d'utilisation.  Utilisez uniquement un couvre-matelas NUVO afin d'éviter les risques de décharges électrostatiques.  Si le cordon d'alimentation est endommagé, remplacez-le immédiatement ou utilisez la source d'alimentation électrique interne de la table jusqu'à...
  • Page 49: Pinch Point Warnings

    B2AV800 Series Surgical Table IM-B2AV800-E01  Si le champ opératoire s'accroche dans les points d'articulation ou sur les couvercles d'articulation (les plaques en acier inoxydable couvrant les articulations), tout mouvement risquerait de causer des dommages à la table. Bordez les draps entre le couvre-matelas et la table afin d'empêcher cela.
  • Page 50 B2AV800 Series Surgical Table IM-B2AV800-E01...
  • Page 51 B2AV800 Series Surgical Table IM-B2AV800-E01 Figure 5-2: Definition of Buttons & Indicators of Hand Pendant 2 and 8 are for models with L suffix. 14 and 21 are for models With S suffix The first six #s are the part number, the last # is version number.
  • Page 52: Table Positioning

    B2AV800 Series Surgical Table IM-B2AV800-E01 Description Description Floor Lock Indicator (Green LED): Lock Battery Level Indicator Floor Lock: Lock Interlock Indicator (Yellow LED) Normal Orientation Indicator (Green LED) Service Indicator (Yellow LED) Normal Patient Orientation AC Power Indicator (Green LED) Reverse Orientation Indicator (Green LED) Reflex Reverse Patient Orientation...
  • Page 53: Level Button Operation

    Service Code. For other events, which the Service LED is lit only while a button is pressed; press the button while looking for the Service Code display. Let go of the button to exit from the Service Code display. Inform your finding to the NUVO 5-11...
  • Page 54 B2AV800 Series Surgical Table IM-B2AV800-E01 Sales or Service Representatives. (Refer to Table 5-1 for description of Service Codes.) Table 5-1 Service Codes Display/Code Symptom Cause Pump motor is not working. 1. Overheated motor 2. Faulty driver PCB (ASM0023) 1. LOCK functions normally but signal Faulty 2-set photo sensors from both of 2-set photo sensors (ASM0026) or connection...
  • Page 55 B2AV800 Series Surgical Table IM-B2AV800-E01 Display/Code Symptom Cause 13 When rotary valve keeps turning but 1. Short circuit within the angle receives no signal from U1~U4 of system. ASM0027 2. Rotary valve motor damaged 3. ASM0023 driver PCB damaged 4. ASM0027 or its cable damaged Rotary valve position number not ASM0027 or its cable damaged...
  • Page 56: Using The Override System For Table Positioning

    B2AV800 Series Surgical Table IM-B2AV800-E01 Display/Code Symptom Cause Voltage is too low (i.t.)* Wrong setting Not used Not used Wrong DIP Switch setting (i.t.)* (it)*: Initiating test of system, which happens when the system is being power up after it has been reset by being disconnected from all power sources. 5.2.2 Using Override System for Table Positioning WARNING!
  • Page 57 B2AV800 Series Surgical Table IM-B2AV800-E01 AVERTISSEMENT! RISQUE D'EXPLOSION: N'utilisez pas la table avec des anesthésiques inflammables. RISQUE DE MOUVEMENTS INATTENDUS DE LA TABLE: Si la table chirurgicale commence à se déplacer par elle-même alors que l'interrupteur de commande prioritaire est en position « haute » ou « basse », placez immédiatement l'interrupteur en position «...
  • Page 58 B2AV800 Series Surgical Table IM-B2AV800-E01 Figure 5-3: Definition of switches and items on the Base Panel and Override System 5-16...
  • Page 59 B2AV800 Series Surgical Table IM-B2AV800-E01 Description Rotary Knob (#5) Override Control (#11) Fuse (neutral) Description Down Fuse (live) Floor Lock Lock Unlock Wire Collect Tie Slide To head To leg Power Cord Socket Height Down Manual Rotary Knob Tilt Left Right Foot Control Socket Trendelenburg...
  • Page 60: Head Section Positioning

    B2AV800 Series Surgical Table IM-B2AV800-E01  When Override System is enabled, Hand Pendant and Foot Controller are disabled, and the Hand Pendant’s Service LED is lit. Remarque:  Le système de commande de priorité ne convient qu'à l'orientation NORMALE du patient. ...
  • Page 61: Kidney Elevator Instructions

    B2AV800 Series Surgical Table IM-B2AV800-E01 To adjust downward: Lift up leg section a little bit with one hand; pull the ring-pull with the other hand to release the ratchet mechanism. Adjust the leg section to a desire angle before letting go of the ring-pull. Move around the leg section to make sure the ratchet mechanism is securely locked.
  • Page 62: Cassette Channels

    B2AV800 Series Surgical Table IM-B2AV800-E01 1. Locate the ratchet wrench (#12, Figure 3-2) under the right side of Back section, pull down and bring out the handle. 2. Adjust the lever to the left for the upward direction, and then use the wrench to bring up the kidney elevator until it reaches the desired height.
  • Page 63: Backup Batteries

    B2AV800 Series Surgical Table IM-B2AV800-E01 WARNING! AVERTISSEMENT! insulation. Carefully read the manufacturers’ instructions of the aforementioned products, as they relate to patient safety. RISQUE DE BLESSURES AU PATIENT: Un danger potentiel est présent lorsque la table chirurgicale est utilisée avec de l'équipement chirurgical de haute fréquence tel que des défibrillateurs cardiaques ou des moniteurs à...
  • Page 64: Accessories

    RISQUE D'INSTABILITÉ: L'utilisation de la table NUVO et de ses accessoires ou d'accessoires fabriqués et vendus par d'autres compagnies pour d'autres fins que celles indiquées peut entraîner des blessures au patient ou à...
  • Page 65: Side Rail Locks And Clark Sockets

    To detach the mattress pad, pull up and peel it away from the tabletop. Note: Use Pads supply by NUVO to ensure its anti-static properties. Remarque: Utilisez les couvres-matelas fournis par NUVO pour garantir les propriétés antistatiques. Part Number Description...
  • Page 66: Arm Board

    B2AV800 Series Surgical Table IM-B2AV800-E01 Many accessories are attached to the table’s side rails; these accessories need side rail locks or Clark sockets to help secure them in place. To install a side rail lock (figure 5-10), place it on the side rail, then insert the rectangular rod of accessory through the corresponding rectangular hole, and lock the accessory into position with thumbscrew.
  • Page 67: Restraint Strap

    5.3.5 Cassette Top An optional accessory of four-section cassette top is available from NUVO for B2AV800 Series table. An X-ray cassette can be inserted between tabletop and the cassette top from the left for right side of table. Align and push the pins of cassette top into holes on the tabletop so they fit tightly, then place the patient onto the table top.
  • Page 68: Foot Controller

    5.3.6 Foot Controller An optional accessory Foot Controller is available from NUVO for B2AV800 Series table. Insert the controller’s signal cord into the Foot Control Socket on the Base Control Panel (#6, Figure 5-3) and then tighten the screw cap to prevent the disconnection.
  • Page 69: Maintenance

    Cleaning and Disinfection Procedures WARNING! AVERTISSEMENT! PERSONAL INJURY HAZARD: Only recommended cleaners/disinfectants should be used on NUVO surgical tables. Do not use alcohol, which does not have sufficient cleaning/disinfecting properties, or phenolics, which when inadequately rinsed off could burn the patient’s skin.
  • Page 70 Heavy and dried-on soil should be pre-soaked. c. DISINFECTION: Follow the product label and dilute disinfectants and/or germicides accordingly. Contact NUVO for a list of recommended disinfectants. When using disinfectants, read and follow label’s recommendations. Note: Remarque: Lors de l'utilisation de désinfectants, lisez et suivez les recommandations...
  • Page 71 B2AV800 Series Surgical Table IM-B2AV800-E01 CAUTION! MISE EN GARDE! Veuillez lire attentivement l'étiquette du flacon aérosol et suivez toutes les instructions et les avertissements.  Keep all cleaning fluid out of electrical receptacles. Gardez tout liquide de nettoyage hors de contact des réceptacles électriques. Remove soil with a disposable cloth and place the used cloth in an appropriate biohazards waste-disposal container.
  • Page 72: End-Of-Day Cleaning Procedures

    B2AV800 Series Surgical Table IM-B2AV800-E01 To remove any remaining residue, wipe cleaned surfaces again with a clean, damp lint- free cloth. d. Repeat steps a, b and c for all surface. Turn the table’s power off with Hand Pendant when finished with cleaning procedure. 6.2.2 End-of-Day Cleaning Procedures Perform cleaning procedures as described in Section 6.2.1.
  • Page 73: Monthly Maintenance

    B2AV800 Series Surgical Table IM-B2AV800-E01 Turn the table’s power off with Hand Pendant when finished with cleaning procedure. Monthly Maintenance Perform all table functions. Observe if all movements are smooth and if there is any noise. Have a qualified technician check the table for anything out of the ordinary. Never allow inexperienced, unqualified technicians to attempt any repair on the table.
  • Page 74: Lubrication Schedule

    B2AV800 Series Surgical Table IM-B2AV800-E01 Lubrication Schedule LUBRICANT TABLE SECTION MINIMUM FREQUENCY Articulation pins – Leg Section Articulation pins – Tabletop Tilting Grease, Articulation pins – Back Section 2 x per year See Table 2-2 Articulation pins – Seat Section Central column –...
  • Page 75: Battery Charging Procedures

    B2AV800 Series Surgical Table IM-B2AV800-E01 MINIMUM FREQUENCY ITEMS 4. TABLE MOVEMENTS (NORMAL AND REVERSE ORIENTATIONS) Seat and Back Sections Lateral Tilt 1× per year Height and Sliding Head Section 5. TABLE RIGIDITY Check both ends of tabletop for any horizontal or vertical play. Check side tilt mechanism for any play.
  • Page 76: Fuse Replacement

    Les batteries ne doivent être expédiées que par voie maritime. Remplacez uniquement la batterie par un modèle recommandé par NUVO. Lorsque le cordon d'alimentation a été débranché pendant plus de dix (10) jours, le boîtier de commande manuelle est ensuite déchargé.
  • Page 77 Just push it in the cap and turn slightly to unlock it and use hand to pull it out. Figure 6-2 Replace blown fuse(s) with new one(s) according to the specifications below. NUVO Item Id Specifications AC Voltage EF0002...
  • Page 78: System Reset

    B2AV800 Series Surgical Table IM-B2AV800-E01 System Reset 1. Unplug AC and the AC LED that is on the base panel should be off, and vice versa. 2. While the AC is plugged in, turn on hand pendant, and its panel should be lit. 3.
  • Page 79 The listed numbers of spare parts should be kept in stock if you have between 1 and 10 tables. When you order parts, please include the part number, description, and quantity for each replacement part. Send your order directly to your nearest NUVO sales representative. Q’ty Notes Item Id Description...
  • Page 80 As an environmentally conscious manufacturer, we have made the mattress pad outer covering using a biodegradable material. Since the following parts have a limited shelf life, order them from your NUVO sales representatives as needed. Mattress pads can be ordered from NUVO or else directly from the following contacts:...
  • Page 81: Circuit Diagrams

    B2AV800 Series Surgical Table IM-B2AV800-E01 Section 8 CIRCUIT DIAGRAMS ASM0023 Figure 8-1: B2AV800 All Series ASM0021 Override Upper +24V JP19 Override Upper 5EHDVC-02P +3.3V Override Lower Override Lower Main1REL RAW3A ManEna photo F1 COL4 Main1REL (ASM0026) Motor COL3 A2501WV2-03P Override Switch Motor1Dir A2501WV2-4P A2501WV2-5P...
  • Page 82 B2AV800 Series Surgical Table IM-B2AV800-E01 This page intentionally left blank...
  • Page 83: Waste Parts Disposal

    B2AV800 Series Surgical Table IM-B2AV800-E01 Section 9 WASTE PARTS DISPOSAL WARNING! AVERTISSEMENT! DISPOSAL HAZARD: A hazardous waste management company with appropriate licenses and permits may be required to handle some of the materials which are contained inside the surgical table. RISQUE LIÉ...
  • Page 84 B2AV800 Series Surgical Table IM-B2AV800-E01 This page intentionally left blank...