No.
Floor Lock Indicator (Green LED): Lock
1.
Floor Lock: Lock
2.
Normal Orientation Indicator (Green LED)
3.
Normal Patient Orientation
4.
Reverse Orientation Indicator (Green LED)
5.
Reverse Patient Orientation
6.
Floor Lock Indicator (Green LED): Unlock
7.
Floor Lock: Unlock
8.
Reverse Trendelenburg
9.
10. Height Up
11. Left Tilt
12. Back Up
13. Leg Up
14. Slide Up (Head end)
15. Flex
Note:
14 and 21 are for models B2AV1000S, B2AV1000SK and B2AV1000SKP only.
Remarque: Les points 14 et 21 ne concernent que les modèles B2AV1000S, B2AV1000SK et
B2AV1000SKP.
5.2
Table Positioning
Make sure the Floor Locks are engaged before reading further.
Use Hand Pendant or an optional Foot Controller to create a desired table position, but Head
Section and Kidney Elevator could only be adjusted manually. In case of a Hand Pendant
malfunction, use the Override System before the problem can be solved.
5.2.1
Hand Pendant Operation
Complete Section 4 before using the Hand Pendant for first surgery.
1. Press desired button(s) to create a desired position. Press one button at a time. There
are no automatic buttons other than the LOCK and UNLOCK buttons, the table moves
only when a movement button is pressed, and it stops as soon as the button is released,
or it stops when it has reached its maximum or programmed allowable distance or angle.
2. Release the button when the desired position is reached.
3. For LOCK or UNLOCK, press the button and release it after the function has been
activated. The table will automatically LOCK or UNLOCK.
Description
No.
Battery Level Indicator
16.
Interlock Indicator (Yellow LED)
17.
Service Indicator (Yellow LED)
18.
AC Power Indicator (Green LED)
19.
Reflex
20.
Slide Down (Leg end)
21.
Leg Down
22.
Back Down
23.
Right Tilt
24.
Height Down
25.
Trendelenburg
26.
Tabletop Level
27.
Sleep Mode
28.
29.
Activate
5-9
B2AV1000 Series Surgical Table
IM-B2AV1000-E01
Description