No.
Description
1.
Fuse (neutral)
2.
Fuse (live)
3.
Wire Collect Tie
4.
Power Cord Socket
5.
Manual Rotary Knob
6.
Foot Control Socket
7.
AC Power Indicator (Green LED)
8.
Battery Level Indicator(10 bars: Full, 5
bars: Mid, 2 bars: Low)
9.
(not used)
10.
ON/OFF Switch (to actuate movement)
11.
Override Control Switch (to control
direction of movement)
5.2.2.2 B2AV1000 / B2AV1000K
The B2AV1000/B2AV1000K tables are equipped with an Override System (Figure 5-4-2) for
controlling table movements in the event of Hand Pendant malfunction. The system consists of
an on/off switch, a manual rotary knob and an override control toggle switch.
1.
Turn knob (# 5, Figure 5-4-2) to select a desired movement.
2.
Toggle switch (# 11, Figure 5-4-2) up or down towards the direction of desired movement.
Note:
If the table moves by itself, return the Override Control toggle switch to its neutral
position, which is in the middle, and call for service.
Remarque: Si la table se déplace d'elle-même, remettez l'interrupteur de commande
prioritaire en position neutre (au milieu), et appelez un technicien qualifié pour la
reparation.
3.
Toggle switch (#10, Figure 5-4-2) to activate the movement.
4.
Repeat step 1, 2 and 3 as many time as needed to create a desired table position.
Note:
Override System is only suitable for NORMAL patient orientation.
Although table positioning can be performed with the Override System, the problem
should be corrected as soon as possible. Have a qualified service technician repair
the table before further use.
Rotary Knob (#5)
No.
Description
12.
Floor Lock
13.
Leg
14.
Height
15.
Tilt
16.
Trendelenburg
17.
Back
5-18
B2AV1000 Series Surgical Table
IM-B2AV1000-E01
Override Control (#11)
Up
Down
Lock
Unlock
Up
Down
Up
Down
Left
Right
Rev. Trend.
Trend.
Up
Down