6.3
Monthly Maintenance
Perform all table functions. Observe if all movements are smooth and if there is any noise.
Have a qualified technician check the table for anything out of the ordinary. Never allow
inexperienced, unqualified technicians to attempt any repair on the table.
6.4
Quarterly Maintenance
SHOCK HAZARD: Monitor the electrically conductive properties of casters and floor
locks. Measure their resistance level according to the requirements of NFPA 99
before placing table in use. Subsequent routine testing should be performed at least
once every three months.
RISQUE DE DÉCHARGE ÉLECTRIQUE: Surveillez les propriétés de conductivité
électrique des dispositifs de verrouillage au sol. Mesurez leur niveau de résistance
selon les exigences de la norme NFPA 99 avant de commencer à utiliser la table.
Des tests de routine doivent être effectués au moins une fois tous les trois mois.
ELECTRIC SHOCK HAZARD: Only authorized service personnel should be allowed to
remove the covers.
RISQUE DE DÉCHARGE ÉLECTRIQUE: Seul le personnel de service autorisé devrait
être autorisé à retirer les couvercles.
1.
Clean the floor locks according to 6.2.2 and then check for conductivity.
WARNING! AVERTISSEMENT!
6-5
B2AV1000 Series Surgical Table
IM-B2AV1000-E01