7. Standards
he Compex is based directly on
T
medical technology.
o guarantee your safety, the design,
T
manufacturing and distribution of
Compex are in conformity with the
requirements of the European
Directive 93/42/CEE.
he device is in conformity with the
T
standard for general safety rules for
electromedical devices IEC 60601-1,
the standard for electromagnetic
compatibility IEC 60601-1-2 and the
standard for special safety rules for
nerve and muscle stimulators IEC
60601-2-10.
urrent international standards (IEC
C
60601-2-10 AM1 2001) require that a
warning be given concerning the
application of electrodes to the thorax
(increased risk of cardiac fibrillation).
irective 2002/96/CEE Waste electrical
D
and electronic equipment (WEEE).
8. Patent
m-4
:
US Patent 6,324,432. Patents pending
in the USA, Japan and Europe.
Snap electrode:
Patent pending
All manuals and user guides at all-guides.com
9. Normalised symbols
the stimulation pulses can exceed 10
mA and 10 V. The information given
in this handbook must be strictly
observed.
T
Functions
On/Off (2 positions, stable)
Waiting or preparation for part of the unit 01-06
Stop (switching off)
10. Technical information
General points
P
battery (NIMH) (4,8 V ≈ 1200 mA/h).
C
recharging the battery bear the
following information:
Type TR503-02-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-E-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
Type TR503-02-A-133A03
Input 90-264 VAC /47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-A-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
Type TR503-02-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 400mA / 6W
Type TR1509-06-U-133A03
Input 90-264 VAC / 47-63Hz / 0.5A max.
Output 9V / 1.4A / 15W
mportant: Under some condi-
I
tions, the effective figure for
he Compex is a class II device unit
with its own internal electric
power, with type BF applied
sections.
aste electrical and electronic
W
equipment (WEEE).
he "On/Off" switch is a multi-
T
function key:
Symbol N° (according to CEI 878)
ower supply: NIMH rechargeable
hargers: The only chargers used for
01-03
01-10
93