14a
Fig. 14
14a The Compex has been charging for 3 minutes
and 8 seconds (a full charge may take from 1h00
to 1h30 hours with the rapid charger delivered
with your device).
14b Charging is in progress and the small battery
symbol is not completely full. When charging is
completed, total charge duration flashes and the
small battery is completely full. The Compex
switches off automatically as soon as you discon-
nect the charger.
9. Problems and solutions
Electrode fault
Fig. 15
15-14a The Compex "bleeps" and alternatively
displays the symbol of a couple of electrodes and
an arrow pointing to the channel where a problem
has been detected. Below, the stimulator has
detected an electrode fault on channel 2. This
message may mean that:
1) there is no electrode connected to this channel;
2) the electrodes are old, worn out and/or the
contact is poor: try using new electrodes;
3) the electrode cable is defective: try it using
another channel. If cable shows a default again,
please replace.
All manuals and user guides at all-guides.com
14b
15a
Recharging error
recharging error may happen during
A
charging. The Compex "bleeps" conti-
nuously and the following screen is
displayed:
Fig. 16
f this message appears, disconnect
I
the charger and the device switches
off automatically. Please contact the
hotline, noting this error code.
f, for any reason, you experience any
I
other trouble with your device, please
contact the hotline mentioned and
approved by Compex Médical SA.
115