Alternating Current - Gima SUPERVEGA BATTERY Mode D'emploi

Aspirateur sur chariot
Masquer les pouces Voir aussi pour SUPERVEGA BATTERY:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

25
Recommended separation distance between portable and mobile radio-communication devices and the monitor
The SUPERVEGA BATTERY SUCTION UNIT surgical aspirator is intended to operate in an electro-magnetic environment
where RF irradiated interferences are under control. The client or operator of the SUPERVEGA BATTERY SUCTION UNIT
device can help prevent electro-magnetic interference by keeping a minimum distance between the portable and mobile
RF communication devices (transmitters) and the SUPERVEGA BATTERY SUCTION UNIT device, as recommended
below, in relation to the radio-communication maximum output power.
Maximum nominal
output power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by the reflection from buildings, objects and people
SYMBOLS
Caution: read instructions (warnings) carefully
Keep in a cool, dry place
Manufacturer
Product code
Medical Device complies
with Directive 93/42/EEC
WEEE disposal
Serial number
Battery
Direct current
~

Alternating current

ON / OFF
Separation distance from the frequency transmitter (m)
150KHz to 80MHz
3,5
√P
d=
V
1
0,12
0,38
1,2
3,8
12
ENGLISH
80MHz to 800MHz
12
√P
d=
E
1
0,12
0,38
1,2
3,8
12
Follow instructions for use
Keep away from sunlight
Date of manufacture
Lot number
Type B applied part
Class II applied
Temperature limit
Atmospheric Pressure limit
Humidity limit
Hz
Mains frequency
800MHz to 2,7GHz
23
√P
d=
E
1
0,23
0,73
2,3
7,3
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

28193