6 Transport, storage and disposal
6.1 Decontamination prior to shipment
Before the secretion aspirator Allegra M30 is passed on, the device
must be properly processed by Heinen + Löwenstein GmbH & Co. KG or by
specialized personnel authorized by Heinen + Löwenstein GmbH & Co. KG
to protect subsequent users. Processing is mandatory in accordance with the
German Medical Device Operator Ordinance (MPBetreibV), German Medical
Devices Act (MPG) and the manufacturer's instructions.
Heinen + Löwenstein GmbH & Co. KG offers its partners and customers fast
and professional processing as well as required testing services.
Before the secretion aspirator Allegra M30 is shipped to
Heinen + Löwenstein GmbH & Co. KG, it must be cleaned and disinfected.
Please follow the instructions in section 4.1 resp. 4.2! Please affix the supplied
"Used Medical Device" label to the shipping carton! Please notify
Heinen + Löwenstein GmbH & Co. KG of the return in advance.
EN
6.2 Storage
74
Store the secretion aspirator Allegra M30 in accordance with the
indications in the technical data (section 7)!
Charge the battery of the secretion aspirator Allegra M30 before
storing the device. This ensures that it is operational at all times.
If the secretion aspirator Allegra M30 is not used for a longer period
of time (approximately 10 months), the battery must be fully recharged again!
6.3 Disposal
•
The components of the device must be disposed of properly at
the end of the product's service life.
•
Ensure that the disposed components are clean and carefully
sorted by material.
•
The housing material is labeled with a material symbol mark and
is fully recyclable.
•
Decontaminate the device and the accessories prior to disposal.
•
According to EU Directives 2012/19/EU Waste Electrical and
Electronic Equipment (WEEE) and 2011/65/EU on the restriction
of the use of certain hazardous substances in electrical and
electronic equipment (RoHS II), the device must not be disposed
of in domestic waste.
•
The device and the accessories may be disposed of via
Heinen + Löwenstein GmbH & Co. KG or the service partner.
•
In relation to proper disposal of potential biohazards, contact your
local authority.
•
Outside the EU: Follow the disposal requirements of your country!
7 Technical data
Model name
Allegra M30
Flow rate*
30 l/min ± 2 l/min (high flow)
(measuring point at
suction tube nozzle)
Vacuum
max. -90 kPa (~ -900 mbar; -675 mmHg)
(high vacuum)
Conversion factor: 1 kPa ~ 7.5 mmHg;
1 kPa ~ 10 mbar
Canister
Reusable secretion canister (1000 ml)
Suction tube
PVC, inner diameter 6 mm, wall thickness
(item number: 100713-2)
2 mm, length 150 cm, vacuum stable up to
-0.9 bar
Nominal voltage of the
In:
AC 100 – 240 V ~ 50-60 Hz/1.5 A
power supply unit
Out: DC 12 V/5.0 A
Maximum load current
5.0 A
Permissible input voltage
12 V
Power consumption at 12 V
60 W
Degree of protection
Type BF
pursuant to IEC 60601-1
Risk classification pursuant to
IIa
93/42/EEC, IX
Protection class pursuant to
II
IEC 60601-1
IP degree of protection
IP22
CE marking
CE0044
Sound emission
without vacuum: approximately 60 dB(A)
Ambient conditions
Transport/Storage
ambient temperature: -25 °C to +70 °C
relative humidity:
up to 90 %, non-condensing
air pressure: 700 hPa to 1060 hPa
Operation
ambient temperature: +5 °C to +40 °C
relative humidity:
15 % to 90 %, non-condensing
air pressure: 700 hPa to 1060 hPa
Battery
10.8 V; 5800 mAh (lithium ion battery)
Charging time if battery is
~ 3 h
empty
Charging time if battery is
~ 2 h
approximately 50 % full
Power supply unit
GTM91099-6015-3.0-T2, GlobTek Inc.
Dimensions basic device
213 mm x 260 mm x 124 mm
(H x W x D)
Weight (basic device)
2.8 kg
Operating time
5 minutes ON; 5 minutes OFF
Runtime in battery
approximately 45 minutes,
operation
depending on the runtime of the motor
Lifetime
5 years
Item number (REF)
1510215hl
* The information provided may differ depending on the altitude above sea level, the
prevailing air pressure and the air temperature.
EN