3 Operation

3.1 Set-up and startup

The following section describes the operating elements, connections and the
startup of the secretion aspirator Allegra M30:

3.1.1 Connecting the secretion aspirator Allegra M30

Known or identifiable conditions for medical care
within a domestic environment
Children and pets should be kept away from the device to ensure
that it cannot be knocked over or dropped.
Prior to connecting the power supply unit, ensure that the voltage
of the device corresponds to the domestic power supply.
The device may not be used in damp rooms, bathrooms or show-
ers. Avoid wet conditions on power supply unit, control panel and
socket for the power supply unit. Never submerge the device in
water or other liquids (also not while it is switched off).
Check the power supply unit incl. mains cable for potential damages
EN
before each use and replace it if there is any damage.
58
Use the socket for the 12 V power supply unit (fig. 1(D)) to connect the device
to the mains power supply with the aid of the supplied power supply unit to
charge or operate it as needed. Make sure that the device is positioned in such
a way that it is possible to easily disconnect it later.
Information on the permissible environmental conditions during operation can
be found in section 7 "Technical data".
fingertip
suction tube
aspiration catheter
device with resuable
patient
secretion canister
power supply unit
(type: GTM91099-
6015-3.0-T2)
mains cable
power outlet
Fig. 2
Connection of the secretion aspirator Allegra M30 to patient and accessories

3.2 Startup

Take the device and the accessories out of the packaging.
Always place the device on a sturdy and flat surface without slope.
Prior to the initial startup it is essential to follow the safety instructions in
section 1.6.
Only operate the device with inserted bacterial filter and connected
secretion canister. Bacterial filter and secretion canister are already fitted on
delivery. Note the following instructions for all further applications.
Always have one backup bacterial filter at hand, since it is absolutely essen-
tial for safe operation!
Prior to each application, check that the reusable secretion canister is com-
pletely clean to avoid frothing.
3 Operation

3.2.1 Connecting the reusable secretion canister

1. Fold the moving part of the secretion canister lid slightly upwards and place
the secretion canister lid on the secretion canister.
Fig. 3
Fig. 4
EN
59