4 Maintenance
•
The secretion aspirator Allegra M30 is maintenance-free if used
in accordance with the instruction for use, exempt from components with
limited durability.
•
Perform a visual and functional check prior to each application. Also include
the accessories of the device in doing so.
•
The device may only be opened and repaired by
Heinen + Löwenstein GmbH & Co. KG or by specialized personnel
authorized by Heinen + Löwenstein GmbH & Co. KG in compliance with
the service documents specified by the manufacturer as well as with techni-
cal and hygienic safety measures.
•
The device may be returned to Heinen + Löwenstein GmbH & Co. KG for
repair either directly or via the specialty dealer you purchased the device
from.
•
All accessories must be cleaned and disinfected before the device is
returned. The device itself may be treated with a surface disinfectant.
•
In order to avoid time lags, notify Heinen + Löwenstein GmbH & Co. KG
prior to returning the device and mark the shipping carton of the device
with the label "Used Medical Device".
Heinen + Löwenstein GmbH & Co. KG neither ensures proper operation
EN
for the secretion aspirator Allegra M30 nor is
Heinen + Löwenstein GmbH & Co. KG liable for any resulting property
72
damage or personal injuries if:
•
original accessories and spare parts are not used,
•
the user instructions in this instruction for use are not followed,
•
installation, settings, modifications, enhancements or repairs are not
performed by Heinen + Löwenstein GmbH & Co. KG or by personnel
authorized by Heinen + Löwenstein GmbH & Co. KG.
4.5 Inspection of the secretion aspirator Allegra M30
Heinen + Löwenstein GmbH & Co. KG offers its partners and custom-
ers fast and professional processing as well as required testing services.
5 Troubleshooting
Malfunction
Display
Probable cause
Device does not
do not illuminate
Battery is discharged.
start
blinks red
Battery is discharged and impermissible input
voltage.
blinks red
Internal error
Device does not
blinks red
•
Battery temperature too high for charging
charge
and
illuminates green
•
Battery is defective.
blinks red
Impermissible input voltage
Insufficient flow
illuminates green/orange
•
Bacterial filter is blocked.
rate
•
Leak in the suction line or in the canister
and/or
illuminates green
•
Service life of vacuum pump expired
No flow rate
continuous red light
•
Float sphere of the overflow protection blocks
the aspiration lid inlet.
•
Bacterial filter is blocked.
•
Suction tube is blocked.
•
Bacterial filter is not inserted.
Device
blinks red
Internal error
switches off
Remedy
Connect the power supply unit for charging.
Connect the correct power supply unit, check
the power supply.
Please contact service!
•
No measure required, automatic resump-
tion of charging when battery has cooled
down.
•
Please contact service!
Connect the correct power supply unit, check
the power supply.
•
Replace the bacterial filter.
•
Check that secretion canister lid and tube
connections are tight and no leakages are
EN
apparent.
•
Please contact service!
73
•
Check fluid level in the secretion canister,
empty the canister.
•
Replace the bacterial filter.
•
Replace the suction tube.
•
Insert the bacterial filter.
Please contact service!