If the top cover is heavily soiled or if penetration of the top cover is suspected,
use the following process for cleaning.
1. Unzip the top cover and remove.
NOTE: The top cover can be laundered using low temperatures up to
<70˚F (21˚C) with chemicals suitable for low temperature washings and
prepared in recommended dilutions. Allow to air dry. Ensure cover is
completely dry before storage.
2. After top cover is removed, wipe down all visible surfaces with the hospital
grade disinfectant mixed according to manufacturer's recommendations.
3. Wipe off excess solution.
4. Wipe entire surface a second time with the hospital grade disinfectant mixed to
manufacturer's recommendations.
5. Rinse the surface thoroughly using plain water.
6. Allow to air dry.
7. Examine the three individual section covers for any tears that may have allowed
contamination to penetrate this layer.
8. If any of these section covers are breached, remove the cover and disinfect the
surface below it using the two-step process noted above.
9. Allow to air dry.
10. Replace any torn covers with a new one.
Repairs
It is recommended that any maintenance or repairs beyond those described in the
guide be performed by ArjoHuntleigh authorized personnel.
Product Specifications**
The RIK Fluid Overlay
Dimensions
Weight
Top cover material
supplemental mattress
Dimensions
Weight
part numbers
Complete assembly
Supplemental mattress
replacement parts
Top cover
End pad
Center pad
Bottom panel
**Specifications subject to change without notice.
78 in L x 33 in W x 5.5 in H (198 cm x 83 cm x 13.9 cm)
99 lb (45 kg)
Nylon fabric with moisture vapor permeable backing
78 in L x 33 in W x 2.5 in H (198 cm x 83 cm x 6.4 cm)
10 lb (4.5 kg)
VL22000E
MS22000
TC2200E
ES2200E
CS2200E
BC22000
8