11.2 Default Alarming Values and Setup Range

The default alarming value:
Parameter
SYS
DIA
MAP
SpO
2
Pulse rate
The high and low limits setting range:
Parameter
SYS
DIA
MAP
SpO
2
Pulse rate
Note: Alarm signal (for any alarm source) generation delay time: <1 sec, refer to the monitor in your hand.
User Manual for Vital Signs Monitor
Mode
Adult
High limit
180mmHg
Low limit
60mmHg
High limit
120mmHg
Low limit
50mmHg
High limit
160mmHg
Low limit
50mmHg
High limit
100%
Low limit
90%
High limit
180bpm
Low limit
40bpm
Mode
Adult
High limit
(30~280) mmHg
Low limit
(29~279) mmHg
High limit
(11~232) mmHg
Low limit
(10~231) mmHg
High limit
(21~242) mmHg
Low limit
(20~241) mmHg
High limit
1~100%
Low limit
0~99%
High limit
(1~300) bpm
Low limit
(0~299) bpm
Pediatric
130mmHg
50mmHg
90mmHg
40mmHg
110mmHg
40mmHg
100%
85%
200bpm
50bpm
Pediatric
(30~200) mmHg
(29~199) mmHg
(11~150) mmHg
(10~149) mmHg
(21~165) mmHg
(20~164) mmHg
1~100%
0~99%
(1~350) bpm
(0~349) bpm
67
Neonate
110mmHg
50mmHg
90mmHg
30mmHg
100mmHg
30mmHg
100%
85%
220bpm
50bpm
Neonate
(30~135) mmHg
(29~134) mmHg
(11~100) mmHg
(10~99)mmHg
(21~110) mmHg
(20~109)mmHg
1~100%
0~99%
(1~350) bpm
(0~349) bpm