Chapter 5 Alarm; Alarm Priority; Alarm Signal Generation - Gima 35132 Manuel De L'utilisateur

Moniteurs de signes vitaux
Masquer les pouces Voir aussi pour 35132:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.1 Alarm Priority

Low Priority:
NIBP over range
Temp over range
ECG Lead off
SpO
Probe off
2
Temp Probe off
NIBP error message
SpO
error message
2
Medium Priority:
HR/PR over range
High Priority:
ECG Arrest
Ventricular Run
Ventricular fibrillation
SpO
over limit
2
SYS over limit
DIA over limit
MAP over limit
Temp over limit
HR/PR over limit
Can not detect SpO
2
Can not detect HR/PR

5.2 Alarm Signal Generation

When there is alarm condition, the monitor generates alarm signal with visual indications (which are shown by
two ways: LED indicator with different color and textual message display) and audible indication.
Visual Alarm Indication
The flashing rates for the three categories of alarms are shown in the table below.
LED Indicator Color
Red flashing
Yellow flashing
Yellow light
Refer to Chapter 11 Alarm Information for detailed alarm message descriptions.
User Manual for Vital Signs Monitor

Chapter 5 Alarm

Alarm Category
High priority alarm
Medium priority alarm
Low priority alarm
Table 5.1
46
Flashing Rate
2 Hz
0.5 Hz
Constant(on)(non-flashing)