NIBP Setup Screen Description:
² "SYS alarm high": high alarm limit setting for systolic pressure.
² "SYS alarm low": low alarm limit setting for systolic pressure
² "DIA alarm high": high alarm limit setting for diastolic pressure.
² "DIA alarm low": low alarm limit setting for diastolic pressure.
² "MAP alarm high": high alarm limit setting for mean arterial pressure.
² "MAP alarm low": low alarm limit setting for mean arterial pressure.
² "PR alarm high": high alarm limit setting for PR. Setting range: 1%~299%, the factory default is "190".
² "PR alarm low": low alarm limit setting for PR. Setting range: 0%~298%, the factory default is "40".
² Initial pressure: Cuff pressure to be inflated initially, its range is different depending on patient type.
for neonates: initial inflation pressure: setting range: 60.~80 mmHg; default value: 70 mmHg.
for pediatrics: initial inflation pressure: setting range: 80~140 mmHg; default value: 120 mmHg.
for adults: initial inflation pressure:setting range: 80~200 mmHg; default value: 150 mmHg.
Note: if the device is configured with "SunTech" blood pressure module, then the initial inflation
pressure setting range for adult is 120~200mmHg.
Note: In order to avoid inappropriate initial pressure value which may cause harm to patients, when
patient type is changed or measuring mode is altered or patient ID is changed, the inflating pressure
value will rollback to the latest setting value.
² "Unit": the pressure unit. mmHg and kPa for optional.
² "NIBP Mode": NIBP measuring mode, "STAT", "Manual", "CUSTOM", "AUTO 1", "AUTO 2", "AUTO 3",
"AUTO 4", "AUTO 5", "AUTO 10", "AUTO 15", "AUTO 20", "AUTO 30", "AUTO 40", "AUTO 50", "AUTO 60",
"AUTO 90", "AUTO 120", "AUTO 240 and"AUTO 480" are optional. When "STAT" is selected, it means the
device will do a short-term (5 minutes) automatic NIBP measurement. "AUTO 1" means NIBP
measurement takes once every minute automatically; "AUTO 480" means NIBP measurement takes
once every 480 minutes automatically; In AUTO mode, the counting-down timer is displayed in MAP
(Time) segment on upper right corner.
Note: when "STAT" (short-term automatic NIBP measurement) mode is selected, the MAP value will
User Manual for Vital Signs Monitor
Figure 4.12 NIBP Setup