Removing Internal Components - natus OLYMPIC SMART SCALE 60 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Section 3: Service
Page 3-16

Removing Internal Components

To remove internal components:
1. Remove the scale enclosure (see Disassembly on page 3-6).
2. The H-shaped tray supporting structure above the loadcell,
called the "spider," must be removed before the PCBs can be 
replaced. Remove the spider from the loadcell by unscrewing
the two mounting screws (Fig. 3N).
Upper Overload Stop
3.  Remove the display PCB by disconnecting the ribbon cable 
from the display board, then unscrew the four mounting screws
(Fig. 3O).
From Control Panel
Mounting Screw (4)
View looks up into enclosure
4. Remove the power supply board by disconnecting all
connectors, then unscrew the mounting screws (Fig. 3K).
OLYMPIC SMART SCALE
Spider
Screws securing
Spider to Loadcell
Overload Stop Screws
Fig. 3N
Display PCB
Ribbon Cable Con-
nector
Fig. 3O
®
Model 60/65
Loadcell

Publicité

Chapitres

Table des Matières
loading

Produits Connexes pour natus OLYMPIC SMART SCALE 60

Ce manuel est également adapté pour:

Olympic smart scale 65

Table des Matières