Sommaire des Matières pour natus OLYMPIC SMART SCALE 60
Page 1
Instruction Manual OLYMPIC SMART SCALE ® Model 60/65 Read and be familiar with this manual before installing, operating, or servicing this device. To ensure operator, CAUTION technician, and patient safety, use only as specified in this manual. 028480 Revision date 9-4-2019...
Page 3
The information in this manual is subject to change without notice. No part of this manual may be photocopied, reproduced, translate, or reduced to any electronic medium without the express written permission of Natus Medical Incorporated. All tradenames and trademarks mentioned herein are property of their respective owners.
Table of Contents Section 1: Installation Intended Use ....................1-1 Model 60 Installation ..................1-1 Calibration Check ..................1-1 Carrying the Scale ..................1-1 Cautions .......................1-1 Locking Units to Grams (Optional) ...............1-2 Instructions for Cart Mounting ..............1-3 Attaching and Removing Tray ..............1-3 Electrical Requirements ................1-4 Electromagnetic Compatibility ..............1-4 Battery Operation ..................1-4 Initial Battery Charge ..................1-4...
Page 6
Table of Contents Setting the Averaging Signal Volume ............2-7 Setting the Brakes on Cart-Mounted Scales ..........2-7 Cleaning .......................2-7 Charging the Battery ..................2-8 Charge Life ....................2-8 Low Battery (LO BAT) ...................2-8 Battery Pack Replacement ................2-8 Section 3: Service Cautions .......................3-1 Service Information ..................3-1 ESD and Contamination ................3-1 Calibration ....................3-1 Calibration Procedure ...................3-2...
Section 1: Installation Cautions CAUTION: Read and be familiar with this instruction manual before using this device. CAUTION: This device should only be assembled or installed by a qualified technician. WARNING: Electrical shock hazard when enclosure is open. Always unplug the power cord from the electrical outlet before assembling, cleaning, or servicing.
Section 1: Installation Locking Units to Grams (Optional) The standard configuration of the scale allows you to toggle the units of measurement between metric and English (grams and pounds). You can permanently lock the units to measure grams only. 1. Press On/Off on the front panel to turn on the scale. 2.
Section 1: Installation Instructions for Cart Mounting Four threaded (10-32 UNF) holes are provided in the base plate of scale for securing it to any cart or table (Fig. 1C). The rubber feet Anchor Holes (4) may remain in place. CAUTION: If using the scale on a cart, ensure that the scale is securely attached to the cart.
Section 1: Installation Electrical Requirements The scales are rated at 120 Vac, 60 Hz, 0.2 amps. A 6-ft. (1.8 m) power cord is included with each scale. CAUTION: Only connect this device to a hospital-grade electrical outlet. Electromagnetic Compatibility This equipment complies with EMC limits for the Medical Device Directive 93/42/EEC (EN 55011Class A and EN 60601-1-2).
Section 1: Installation Shipping/Storage Environment Temperature: -40ºF to 158ºF (-40ºC to 70ºC) Storage Temperature: -4ºF to 113ºF (-20ºC to 45ºC) Humidity: 10% to 100% relative humidity Pressure: 0.5 to 1.0 atmospheres Symbols The following international safety symbols are used on the scale: Table 1 Symbols Symbol...
Section 2: Operation Cautions CAUTION: Read and be familiar with this instruction manual before using or servicing the scale. CAUTION: This product must be connected directly to a hospital-grade electrical outlet. CAUTION: Tray must be securely fastened to its base before use.
Section 2: Operation Fig. 2A Controls 1 On/Off: Turns scale display on/off. (If 4 Recall: If scale is off, press to display last operating on batteries, the scale turns itself off weight. automatically after five minutes of inactivity.) 5 Gm/lbs: Press to convert grams to Zero: Press to zero (cancel or tare) blankets pounds-ounces and vice versa (ounces and other objects.
Section 2: Operation Fig. 2B Status Lights 1 STANDBY: When flashing, the scale is 5 5 gm: The scale is set to weigh within five- warming up and automatically zeroing. Do not grams resolution. touch tray; wait until the “Standby” light dims 6 OVERLOAD: There is too much weight on the and “0”...
Section 2: Operation On/Off Press On/Off to turn the Smart Scale display on or off, respectively. When the Smart Scale display is on, the “Standby” light blinks for a few seconds while the scale warms up and automatically zeros. NOTE: Do not touch tray until the “Standby” light dims. The “Standby”...
Section 2: Operation Weighing Tips: Remember, if an object is on the tray and the display shows “0” the weight of the object has been cancelled (tared out). If an additional object, such as a baby, is then placed on the tray, only its weight will be shown.
Section 2: Operation Weight Lock The weight remains locked on the display until Zero is pressed or scale display is turned off. (When operating on batteries, the display automatically turns off after five minutes of inactivity.) Reweigh After the baby has been weighed (“Read” light illuminates), the baby can be reweighed by pressing Reweigh.
Section 2: Operation Overload If the load exceeds 15 kg (33 lbs-1oz), the overload signal on the display panel will illuminate. Reduce the weight on the scale. Setting the Averaging Signal Volume The Smart Scale sounds an audible beep to alert users when it is computing a weight in “Averaging”...
Section 2: Operation Charging the Battery The Smart Scale has a rechargeable battery pack that automatically charges (“Charging” signal illuminates on the display panel) whenever the scale is connected to an electrical outlet. NOTE: The scale can be used while connected to an electrical outlet with the battery pack removed, or while the batteries are charging.
The Olympic Smart Scale is designed to allow for performance of service in the field to the printed circuit board (PCB) level. These instructions will assist you in isolating the problem PCB or component. After testing, if a PCB or the loadcell is identified as damaged, return the scale to Natus Medical for repair. To identify the problem, see Troubleshooting on page 3-20. ESD and Contamination The integrated circuits and semiconductors on the PCBs can be damaged by electrostatic discharge or contamination by body oil.
Section 3: Service Calibrate Before First Use: The Smart Scale was calibrated at the factory in Seattle prior to shipment. Due to movement during shipment and geographic variations in the force of gravity (which can affect calibration by as much as 20 gm), calibrate the scale before initial use. Calibration Frequency: Calibrate the Smart Scale once every 12 months, at a minimum, or whenever a tray is changed.
3. Repeat step 2 for the each test point shown in Fig. 3E. 4. If any of the weights from the test points are ±6 grams different than the displayed weight from the center of the tray, the loadcell may be damaged and require replacement. Contact the Natus Medical Service Department for assistance. Page 3-3 OLYMPIC SMART SCALE ®...
Reorient or relocate the receiving device. Increase the separation between the equipment. Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected. Contact the Natus Medical Service Department. Page 3-4 OLYMPIC SMART SCALE ® Model 60/65...
Disassembly on page 3-6 and then follow the instructions in Problems and Solutions on page 3-10. If the loadcell or main PCB is damaged, return the scale to Natus Medical for service. Natus Medical does not recommend that customers attempt to repair loadcells, and will not warrant customer attempts to repair loadcells.
Control Knob clears the front of the skirt. Next, lift the scale up, out of the base. NOTE: If a cart-mounted model needs to be returned to Natus Medical for servicing, only the scale portion needs to be removed and returned. Do not return the infant tray. Scale...
Page 26
Section 3: Service 4. Carefully slide the front edge of scale over edge of a bench. Remove the front screw and then loosen the two screws on the front bottom hatch (Fig. 3H). Loosen Two Screws Remove Front Screw Front Bottom Hatch Fig.
Page 27
Section 3: Service 8. Internal components are now exposed to allow troubleshooting measurements (Fig. 3K); see Problems and Solutions on page 3-10. CAUTION: To avoid risk of electrical shock when the device is disassembled, only use battery power, as required, for troubleshooting.
Page 28
Section 3: Service Fig. 3K Page 3-9 OLYMPIC SMART SCALE ® Model 60/65...
2. Use appropriate methods to determine if voltage is present at the electrical outlet. 3. Ensure that all fuses and connectors are properly seated. 4. Check the supply voltages on main and power supply PCBs as shown in Fig. 3L. If any voltages are out of tolerance, contact Natus Medical about replacing the defective PCB. Test Point Location PCB Value J4-1 Power supply 12.00 ±0.25Vdc...
Page 30
(see page 3-13). The displayed counts must be between 500–5000. If out of range: a. Test the loadcell (see page 3-13). If damaged, return the scale to Natus Medical for repair. b. Return to Natus Medical to service the main PCB. 2. If the displayed counts fluctuate more than ±4 counts: a. Ensure that the scale is on a stable, non-vibrating surface and is protected from moving air.
Page 31
3. Ensure that pressing the inoperative switch pulls its signal line to ground on the main PCB. a. If not, test display PCB ribbon cable. Replace if non-functional. b. If so, return to Natus Medical to service the main PCB. Testing the display Simultaneously press gm/lb while powering on the scale. All LEDs illuminate with the exception of “Charging,” which will only illuminate if line power is present.
Page 32
32768 ±10 counts (see Adjusting the Zero Offset on page 3-17). If step 2 is not possible, return the scale to Natus Medical for loadcell replacement. Weight Lock Inhibit Mode Pressing gm/lb twice enters Weight Lock Inhibit Mode; in this mode, weights cannot lock on the display.
The battery pack is ® rated at 7.2 Vav, 1700 mAh, rapid charge. NOTE: Only replace the battery pack with Natus battery pack, catalog no. 400850-01. WARNING: Use of incorrect battery pack could lead to equipment damage, explosion, or fire.
2. Remove the scale enclosure (see Disassembly on page 3-6). 3. Remove fuse from fuse holder (Fig. 3K) and replace with 1/4 amp, 250 Vac Slo-Blo fuse, Natus Medical catalog no. 200021-581. 4. Reassemble the scale by reversing the steps described in Disassembly on page 3-6.
Section 3: Service Removing Internal Components To remove internal components: 1. Remove the scale enclosure (see Disassembly on page 3-6). 2. The H-shaped tray supporting structure above the loadcell, called the “spider,” must be removed before the PCBs can be replaced. Remove the spider from the loadcell by unscrewing the two mounting screws (Fig. 3N). Spider Screws securing Spider to Loadcell...
(On/Off). Raw AD counts are shown. 3. Ensure that the tray is on the scale, then place a 7500-gram weight on the tray. (The weight is available from Natus Medical, catalog no. 56315). 4. Locate the potentiometer port for VR4 (see label on underside of scale).
Section 3: Service Replacing the Tray Mounts If a tray mount becomes stripped and will no longer tighten on the tray leg, replace the mount as follows: 1. Power off the scale and unplug its power cord from the Access Holes (4) electrical outlet. 2. Remove the scale enclosure (see Disassembly on page 3-6). 3.
Printed Circuit Boards and Components When returning equipment for service: 1. Contact Natus Technical Service for a Return Merchandise Authorization (RMA) number and the location where the equipment should be sent. 2. Clean the device, securely package it, and include the RMA number on the outside of the box.
Section 3: Service Troubleshooting The following is a chart of minor problems and their probable solutions. For problems beyond the scope of this chart, call Natus Medical toll free at 1-866-940-7143 in U.S. or Canada. Probable Cause Solution Problem “Standby” light blinks but scale Unit is out of calibration.
Page 40
Section 3: Service Probable Cause Solution Problem “Charging” light does not Replace the fuse (see Replacing the Fuse on Blown fuse. illuminate when power is present. page 3-15). Check the battery pack connection (see Replacing Battery pack is not connected properly. the Battery Pack on page 3-14). Replace battery pack (see Replacing the Battery Battery pack is defective.
Page 49
Mode d’emploi OLYMPIC SMART SCALE ® Modèle 60/65 Lisez ce manuel et familiarisez-vous avec ce dernier avant d’installer, d’utiliser ou d’entretenir cet appareil. Pour assurer la sécurité de l’opérateur, du technicien et du patient, utilisez MISE EN uniquement comme il est indiqué dans ce manuel. GARDE 028480 Date de révision 04-09-2019...
Page 51
Aucune partie de ce manuel ne peut être photocopiée, reproduite, traduite ou réduite en vertu de tout moyen électronique sans la permission écrite expresse de Natus Medical Incorporated. Tous les noms et marques de commerce mentionnés aux présentes appartiennent à leurs propriétaires respectifs.
Page 53
Table des matières Section 1 : Installation Usage prévu ......................1-1 Installation du modèle 60 ..................1-1 Vérification de l’étalonnage ..................1-1 Transporter la balance ....................1-1 Mises en garde ......................1-1 Verrouillage des unités aux grammes (en option) ............1-2 Directives pour le montage sur un chariot ..............1-3 Fixation et retrait du plateau ..................1-3 Exigences électriques ....................1-4 Compatibilité...
Page 54
Table des matières Réglage du volume du signal du calcul de la moyenne ...........2-7 Réglage des freins des balances installées sur un chariot ........2-7 Nettoyage .........................2-7 Chargement des piles ....................2-8 Durée du chargement ....................2-8 Piles faibles (LO BAT) ....................2-8 Remplacement du bloc-piles ..................2-8 Section 3 : Service Mises en garde ......................3-1 Renseignements relatifs au service ................3-1...
Section 1 : Installation Mises en garde MISE EN GARDE : Lisez et familiarisez-vous avec ce mode d’emploi avant d’utiliser cet appareil. MISE EN GARDE : Cet appareil ne doit être assemblé et installé que par un technicien qualifié. AVERTISSEMENT : Danger de choc électrique lorsque l’enceinte est ouverte.
Section 1 : Installation Verrouillage des unités aux grammes (en option) La configuration standard de la balance vous permet de basculer les unités de mesure entre le système métrique et impérial (grammes et livres). Vous pouvez verrouiller de façon permanente les unités de mesure aux grammes seulement. 1.
Section 1 : Installation Directives pour le montage sur un chariot Quatre trous filetés (10-32 UNF) se trouvent dans la plaque de base de la balance pour la fixer sur un chariot ou une table (Fig. 1C). Les pieds Trous d’ancrage (4) en caoutchouc peuvent rester en place.
Section 1 : Installation Exigences électriques Les balances sont classées à 120 V c.a., 60 Hz, 0,2 ampère. Un cordon d’alimentation de 1,8 m (6 pi) est compris avec chaque balance. MISE EN GARDE : Ne reliez cet appareil qu’à une prise électrique de qualité...
Section 2 : Fonctionnement Mises en garde MISE EN GARDE : Lisez et familiarisez-vous avec ce mode d’emploi avant d’utiliser ou d’entretenir cette balance. MISE EN GARDE : Ce produit doit être relié directement à une prise électrique de qualité hospitalière. MISE EN GARDE : Le plateau doit être fixé...
Section 2 : Fonctionnement Fig. 2A Commandes On/Off : Allume ou éteint l’écran de la balance. (En Recall : Si la balance est éteinte, appuyez sur cette cas d’alimentation par des piles, la balance s’éteint touche pour afficher le dernier poids. automatiquement après un délai de cinq minutes.) Gm/lbs : Appuyez sur cette touche pour convertir Zero : Appuyez sur cette touche pour remettre à...
Section 2 : Fonctionnement Fig. 2B Voyants de statut STANDBY : Lorsque ce voyant clignote, la balance 5 5 gm : La balance est réglée pour effectuer la pesée se réchauffe et se remet automatiquement à zéro. à plus ou moins cinq grammes. Ne touchez pas au plateau;...
Section 2 : Fonctionnement Marche/arrêt Appuyez sur la touche On/Off (marche/arrêt) pour allumer ou éteindre l’écran Smart Scale, respectivement. Lorsque l’écran Smart Scale est allumé, le voyant « Standby » (attente) clignote pendant quelques secondes tandis que la balance se réchauffe et se met automatiquement à...
Section 2 : Fonctionnement Conseils de pesée : N’oubliez pas que si un objet se trouve sur le plateau et que l’écran affiche « 0 », le poids de l’objet a été annulé (taré). Si un autre objet, comme un bébé, est posé sur le plateau, seul son poids sera affiché. Si vous devez peser un bébé...
Section 2 : Fonctionnement Verrouillage du poids Le poids demeure verrouillé sur l’écran jusqu’à ce que la touche Zero (zéro) soit enfoncée ou que l’écran de la balance soit éteint. (En cas d’alimentation par les piles, l’écran s’éteint automatiquement après un délai de cinq minutes.) Nouvelle pesée Après avoir pesé...
Section 2 : Fonctionnement Surcharge Si la charge dépasse 15 kg (33 lb 1 once), le voyant « Overload » (surcharge) s'illumine sur l’écran. Réduisez le poids se trouvant sur la balance. Réglage du volume du signal du calcul de la moyenne La balance Smart Scale émet un bip audible pour avertir les utilisateurs qu’elle effectue une pesée en mode «...
Section 2 : Fonctionnement Chargement des piles La balance Smart Scale comprend un bloc-piles rechargeable qui se recharge automatiquement (le signal « Charging » (chargement) s’illumine sur l’écran) lorsqu’elle est branchée sur une prise de courant. REMARQUE : La balance peut être utilisée tandis qu’elle est branchée sur une prise de courant avec le bloc-piles retiré, ou pendant que les piles se rechargent.
Après avoir effectué l’essai, si vous déterminez qu’une carte de circuits imprimés ou que la cellule de mesure sont endommagées, retournez la balance à Natus Medical pour la faire réparer. Pour identifier le problème, reportez-vous à Dépannage à la page 3-20.
Section 3 : Service Étalonnage avant la première utilisation : La balance Smart Scale a été étalonnée à l’usine à Seattle avant son expédition. En raison du mouvement durant l’expédition et des variations géographiques de la force de gravité (qui peuvent affecter l’étalonnage d’une valeur pouvant atteindre 20 g), étalonnez la balance avant de l’utiliser pour la première fois.
6 grammes du poids affiché lors de l’essai au centre du plateau, il est possible que la cellule de mesure soit endommagée et qu’elle doive être remplacée. Communiquez avec le service après-vente de Natus Medical pour obtenir de l’aide. Page 3-3 OLYMPIC SMART SCALE ®...
Augmenter la distance entre l’équipement. Relier l’équipement sur une prise se trouvant sur un circuit différent de celui sur lequel l’autre (les autres) appareil(s) est (sont) relié(s). Communiquer avec le service après-vente de Natus Medical. Page 3-4 OLYMPIC SMART SCALE ®...
3-10. Si la cellule de mesure ou la carte de circuits imprimés principale sont endommagées, retournez la balance à Natus Medical pour la faire réparer. Natus Medical ne recommande pas aux clients de tenter de réparer les cellules de mesure et la garantie ne couvrira pas les tentatives de réparation des cellules de mesure effectuées par les clients.
Ensuite, soulevez la balance vers le haut, hors de la base. REMARQUE : Si un modèle installé sur un chariot a besoin d’être retourné à Natus Medical pour une réparation, seule la partie balance doit être retirée et expédiée. Ne retournez pas le plateau à...
Page 74
Section 3 : Service 4. Glissez soigneusement le rebord avant de la balance sur le rebord d’un banc. Retirez la vis avant, puis desserrez les deux vis situées sur la trappe inférieure avant (Fig. 3H). Déserrer deux vis Retirer la vis avant Trappe inférieure avant Fig.
Page 75
Section 3 : Service 8. Les composants internes sont maintenant exposés afin de vous permettre d’effectuer le dépannage (Fig. 3K); reportez-vous à la section Problèmes et solutions à la page 3-10. MISE EN GARDE : Pour éviter tout risque de choc électrique lors du démontage de l’appareil, utilisez uniquement l’alimentation par piles, au besoin, pour le dépannage.
4. Vérifiez les tensions d’alimentation des cartes de circuits imprimés principale et d’alimentation comme il est indiqué dans la Fig. 3L. Si l’une ou l’autre des tensions se trouve en dehors des tolérances, communiquez avec Natus Medical pour faire remplacer la carte de circuits imprimés défectueuse. Emplacement...
Page 78
Vérifiez la cellule de mesure (reportez-vous à la page 3-13). Si elle est endommagée, retournez la balance à Natus Medical pour la faire réparer. b. Retournez la balance à Natus Medical pour faire réparer la carte de circuits imprimés principale.
Page 79
Sinon, vérifiez le câble ruban de la carte de circuits imprimés d’affichage. Remplacez-le s’il est défectueux. b. Le cas échéant, retournez la balance à Natus Medical pour faire réparer la carte de circuits imprimés principale. Vérification de l’affichage Enfoncez simultanément la touche gm/lb et la touche de mise en...
Page 80
à la page 3-17). Si vous ne pouvez pas effectuer l’étape 2, retournez la balance à Natus Medical pour faire remplacer la cellule de mesure. Mode inhibiteur de verrouillage du poids Appuyez sur gm/lb deux fois pour accéder au mode inhibiteur de verrouillage du poids;...
® de 7,2 V c.a., 1 700 mAh avec chargement rapide. REMARQUE : Remplacez uniquement le bloc-piles par un bloc-piles Natus, nº de catalogue 400850-01. AVERTISSEMENT : L’utilisation d’un bloc-piles inapproprié pourrait causer des dommages à l’équipement, une explosion ou un incendie.
3. Retirez le fusible du porte-fusible (Fig. 3K) et remplacez-le par un fusible Slo-Blo de 250 V c.a., 1/4 ampère, nº de catalogue 200021-581 de Natus Medical. 4. Remontez la balance en inversant les étapes décrites dans Démontage à la page 3-6.
Section 3 : Service Retrait des composants internes Pour retirer des composants internes : 1. Retirez l’enceinte de la balance (reportez-vous à Démontage à la page 3- 6). 2. Le plateau en forme de H soutenant la structure au-dessus de la cellule de mesure, appelé...
3. Assurez-vous que le plateau est sur la balance, puis placez un poids de 7 500 grammes dans le plateau. (Ce poids est offert par Natus Medical, nº de catalogue 56315.) 4. Repérez le port du potentiomètre pour VR4 (reportez-vous à...
Section 3 : Service Remplacement des montures de plateau Si une des montures du plateau est dénudée et ne peut plus être serrée sur le pied du plateau, remplacez la monture comme suit : 1. Éteignez la balance et débranchez son cordon d’alimentation de la prise de courant.
Lorsque vous devez retourner l’équipement pour le faire réparer : 1. Communiquez avec le service technique de Natus pour obtenir un numéro d’autorisation de retour (NAR) et l’endroit où envoyer l’équipement. 2. Nettoyez l’appareil, emballez-le soigneusement et notez le NAR sur la boîte. Remarque : Ne retournez pas le plateau.
Section 3 : Service Dépannage Vous trouverez ci-dessous un tableau décrivant des problèmes mineurs et leurs solutions probables. Pour tout problème non compris dans ce tableau, appelez Natus Medical sans frais au 1-866-940-7143 aux É.-U, ou au Canada. Cause probable Solution Problème...
Page 88
Section 3 : Service Cause probable Solution Problème Le voyant « Charging » (chargement) Remplacez le fusible (reportez-vous à Remplacement des Fusible grillé. ne s’illumine pas lorsqu’il y a du fusibles à la page 3-15). courant. Vérifiez la connexion du bloc-piles (reportez-vous à Le bloc-piles n’est pas bien relié.