Section 3: Service
Page 3-6
Disassembly
WARNING: Electrical shock hazard when enclosure is
removed. Always disconnect the power cord from the
electrical outlet before cleaning or servicing.
CAUTION: PC boards contain static-sensitive parts. Always
use appropriate ESD protection when enclosure is removed.
To disassemble the scale:
1. Turn the scale off and unplug its power cord from the electrical
outlet.
2. Remove the tray by unscrewing four Phillips screws that hold
the tray legs in place (Fig. 3F).
Tray Mount
3. For cart-mounted scales, open the cart door and remove the
four screws that secure the scale to the base (Fig. 3G). Lift
the back of scale and slide scale backwards until the Volume
Control Knob clears the front of the skirt. Next, lift the scale up,
out of the base.
NOTE: If a cart-mounted model needs to be returned to
Natus Medical for servicing, only the scale portion needs
to be removed and returned. Do not return the infant tray.
Volume
Control
Knob
OLYMPIC SMART SCALE
Tray Leg
Phillips screw
Fig. 3F
Scale
Skirt
Base
Fig. 3G
®
Model 60/65