Section 3: Service
Page 3-10

Problems and Solutions

Schematics, wiring diagrams and PCB drawings are located at the 
back of this manual.
Use the troubleshooting measurements, below, to test printed
circuit boards (PCBs) and components that are suspect of 
being defective. For additional information, see Troubleshooting
on page 3-20.
NOTE: The scale will not zero unless the tray is in place.
Ensure that the tray is in place when troubleshooting.
Before performing troubleshooting measurements:
1. Disassemble the scale (see Disassembly on page 3-6).
CAUTION: To avoid risk of electrical shock when the
device is disassembled, use battery power, as required,
for troubleshooting.
2. Use appropriate methods to determine if voltage is present at
the electrical outlet.
3. Ensure that all fuses and connectors are properly seated.
4.  Check the supply voltages on main and power supply PCBs as 
shown in Fig. 3L. If any voltages are out of tolerance, contact
Natus Medical about replacing the defective PCB.
ID
Test Point
Vin
CHG/AC_ON
GND
Vbat
+5Vdigital
+5Vanalog
-5Vanalog
OLYMPIC SMART SCALE
Location PCB
Value
J4-1
Power supply
12.00 ±0.25Vdc
J3-1
Power supply
5.00 ±0.25Vdc
TP1
Power supply
reference
TP2
Power supply
7.2Vdc nominal
TP1
Main
5.00 ±0.1Vdc
TP2
Main
5.00 ±0.1Vdc
TP3
Main
-5.00 ±0.5Vdc
Fig. 3L
®
Model 60/65