Before Attempting Repair - natus OLYMPIC SMART SCALE 60 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Before Attempting Repair
CAUTION: Service should only be performed by a
qualified technician.
WARNING: Electrical shock hazard when enclosure is
removed. Always unplug the power cord from the electrical
outlet before cleaning or servicing.
CAUTION: PC boards contain static-sensitive parts.
Always use appropriate ESD protection when enclosure
is removed.
NOTE: After making any repairs the scale must be
calibrated; ensure that you have the appropriate 7500-gram
weight that is required for calibration before disassembling
the scale. Other repairs involve the use of specific additional 
weights; review the repair information before disassembling
the scale.
To determine the most likely cause of a problem, see
Troubleshooting on page 3-20.
If an internal component needs to be tested, remove the scale
enclosure according to Disassembly on page 3-6 and then follow
the instructions in Problems and Solutions on page 3-10.
If the loadcell or main PCB is damaged, return the scale to 
Natus Medical for service. Natus Medical does not recommend
that customers attempt to repair loadcells, and will not warrant
customer attempts to repair loadcells.
After identifying a problem PCB, see Removing Internal
Components on page 3-16.
NOTE:  Do not remove PCBs for troubleshooting. 
OLYMPIC SMART SCALE
Section 3: Service
®
Model 60/65
Page 3-5

Publicité

Chapitres

Table des Matières
loading

Produits Connexes pour natus OLYMPIC SMART SCALE 60

Ce manuel est également adapté pour:

Olympic smart scale 65

Table des Matières