3. When "C 7500" appears, place the 7500-gram test weight in
the center of the tray. Calibration is successful when "7500"
locks on the display and the "Grams" signal and green "Read"
light illuminate (Fig. 3D).
4. Proceed to the Off Center Load Test.
Off-Center Load Test
To check the accuracy of the load cell after calibration:
1. Place the 7500-gram test weight in the center of the tray and
press Reweigh. Record the weight shown.
2. Place the test weight at one of the four test points shown in
Fig. 3E. Press Reweigh and record the weight shown.
Test point: weight
against side of
tray
Test point: center
of weight 5.5" off
center of tray
3. Repeat step 2 for the each test point shown in Fig. 3E.
4. If any of the weights from the test points are ±6 grams
different than the displayed weight from the center of the tray,
the loadcell may be damaged and require replacement.
Contact the Natus Medical Service Department for assistance.
Fig. 3D
Test point: center of
weight 5.5" off center
of tray
Test point: weight
against side of tray
Fig. 3E
OLYMPIC SMART SCALE
Section 3: Service
®
Model 60/65
Page 3-3