Accessories

If the maximum weight for any of the accessori-
es, e.g. slings, lift bars and any other accessories
is NOT the same as for the lift, then the lowest
max. weight given for any one product is the
one that shall apply. Always check the label on
the lift and lifting accessories or contact Human
Care if you have any questions or if anything is
unclear.

Troubleshooting

Description
Nothing happens when you press the
Up button or the Down button on the
hand control or the control panel
Nothing happens when you press the
Up button or the Down button on the
hand control
Nothing happens when you press the
Up button or the Down button on the
hand control or the control panel.
Nothing happens when you press the
Up button or the Down button on the
hand control or the control panel.
The lift goes in the wrong direction
when you press the Up button or the
Down button on the hand control or
the control panel.
Nothing happens when you press the
Up button on the hand control or the
control panel.
Nothing happens when you press the
Down button on the hand control or
the control panel.
The belt is not fed out when the lift is
on a surface.
The indicator switches from flashing
green or yellow to flashing red.
36
www.humancaregroup.com
Human Care recommends using only Human
Care slings with Human Care lifts. Visit www.
humancaregroup.com to view all our lift acces-
sories.
Indicator
Action
Not lit
1) Check that the mains switch is on.
2) The batteries are fully discharged. Charge the lift
immediately
3) The emergency stop is pushed in. Deactivate by
turning the red button clockwise.
Flashing green
Check the hand control. If the hand control is
Flashing yellow
defective it needs to be replaced. Contact a service
Flashing red
technician.
Constant red
1) Lift motor is overheated. Wait a while and try
again.
2) Lift is overloaded. Lower using the emergency
lowering button
Constant yellow
Charging in progress. Interrupt charging.
Flashing green
The lever (10) has been turned in the wrong direc-
Constant green
tion.
Flashing yellow
Turn the lever towards the connected belt.
Flashing red
Flashing green
The lift is in its maximum upper position. Press the
Flashing yellow
Down button on the hand control or the control
Flashing red
panel to lower the lift.
Flashing green
The lift is in its maximum lower position. Press the Up
Flashing yellow
button on the hand control or the control panel to
Flashing red
raise the lift.
Flashing green
The lift belt has to be pulled upwards at the same
Constant green
time as the Down button on the hand control or the
Flashing yellow
control panel is held pressed in.
Flashing red
Flashing red
The batteries need charging but you can still lift the
client a few times. The lift should be set to charge as
soon as possible.