Roomer S
Guidance and manufacturer's declaration
Human Care lifts mentioned in this document covers following models:
Guidance and manufacturer's declaration – electromagnetic emissions
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should
assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions IEC
61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
+/- 6 kV contact
discharge (ESD)
+/- 8 kV air
IEC 61000-4-2
Electrical fast trans-
+/- 2 kV for power supply
ient / Burst
lines
IEC 61000-4-4
+/- 1 kV for input/output
lines
Surge
+/- 1 kV differential mode
IEC 61000-4-5
+/- 2 kV common mode
Voltage dips, short
<5 % UT
interruptions
(>95 % dip in UT)
and voltage
for 0,5 cycle
variations on power
40 % UT
supply input lines
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT))
for 5 sec
Power frequency
3 A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration – electromagnetic immunity
The Human Care lift is intended for use in the electromagnetic environment specified below. The customer or the user of the Human Care lift should
assure that it is used in such an environment.
38
www.humancaregroup.com
Compliance
Group 1
Class B
Class A
Complies
Compliance level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV for power
supply lines
n/a. for input/output
lines
+/- 1 kV differential
mode
n/a. for common mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT))
for 5 sec
3 A/m
Electro -
magnetic environment - guidance
The Human Care lift uses RF energy only for its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
The Human Care lift is suitable for use in all establishments including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Electro - magnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Human Care lift requires continued ope-
ration during power mains interruptions, it is recommended that the lift
be powered from an uninterruptible power supply or battery.
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment