Troubleshooting

Errors

8.2
Errors
NOTICE!
Property damage due to ignoring error messages!
In the event of an error message, any further use of the GuardOne
the sample material.
- Contact your STARLAB service technician for further assistance.
- Contact details may be found by pressing STARLAB Service & Order Information on the status page.
Fig. 55: Errors
8.3

Identifying non-automatically monitored malfunctions

UV light does not work
In the case of a malfunctioning UV lamp, check the correct fitting of the magnetic switch cable.
1.
Open the front door of the top housing.
2.
Check the connection of the magnetic switch cable.
3.
Make sure that the magnetic switch cable is connected.
4.
Close the front door carefully.
Make sure that the magnetic switch cable does not get caught between the front and the side flap of the top housing.
5.
Enter the settings for the UV lamp again.
In case the UV lamp is still not working, contact your local STARLAB sales representative or STARLAB service technician. For contact details,
press STARLAB Service & Order Information on the status page.
Behaviour in case of a malfunctioning engine
If you hear abnormal engine running noise over a long period of time, there may be a malfunction of the engine.
1.
Turn the main power switch to OFF.
2.
Wait a few minutes before turning the GuardOne
3.
Turn the main power switch to ON.
If the noise of the engine persists, contact your local STARLAB sales representative or STARLAB service technician. For contact details,
press STARLAB Service & Order Information on the status page.
46
· GuardOne
· Manual
®
®
Workstation may result in material damage to the device and
Status errors inform about property damage that prevents the
continued operation of the GuardOne
by a STARLAB service technician.
Blower breakdown (Error 1)
- Blower breakdown means that the GuardOne
detected a malfunction of the blower motor.
Backpressure sensor failure occurred (Error 2)
- Backpressure Sensor failure occurred means that the GuardOne
Workstation has detected a malfunction of the airflow control.
CPU memory block failure occurred (Error 3)
- CPU memory block failure occurred means that the GuardOne
Workstation has detected a malfunction of the device memory.
Behaviour in case of a status error
1.
Contact your STARLAB service technician for further assistance.
2.
For contact details, press STARLAB Service & Order Information on
the status page.
Workstation on again.
®
®
Workstation until repair
Workstation has
®
®
®