Testing the Airflow
1.
In the Home menu, press the Airflow & Light button.
The page for adjusting blower speed or LED light opens.
2.
To adjust the airflow, press onto the percentage value (1) in the Air Flow
field.
3.
Select t he desired airflow rate using the + and - buttons
Selecting an airflow rate of 0% will switch off the blower. To switch it
on again, select an airflow rate above 0%. The available steps are
0, 50, 80 and 100%.
4.
Apply the selection using Apply Settings (3).
If the front door is closed, the airflow will not work. The airflow is wor-
king if the lowest segment of the front door or the whole front door is
open. Even if the magnetic switch is installed improperly, LED light and
blower function will be functional.
When you turn on the airflow, the motor needs about ten seconds until
the airflow starts to blow. When you turn off the airflow, it stops as
soon as the front door is closed.
Testing the LED light
›
You are in the Airflow & Light menu.
1.
To switch on the LED light, use the LED light switch
Your selection is applied immediately.
2.
To select an LED light colour, press LED colour (2). Select the desired
colour, normal white or red, by pressing the + and - button
Your selection is applied immediately.
If the LED lights are not working, please contact your local STARLAB
sales representative.
3.
After you have made sure that the LED lights are working, switch off
the LED light using the LED light switch
4.
To return to the Home menu, either use Back or Home in the
navigation bar.
(2).
(1).
(3).
(1).
Fig. 22: Testing the Airflow
Fig. 23: LED light