Cleaning and Maintenance

Cleaning and Decontamination

Fig. 52: Activate the automatic life time surveillance
7.3
Cleaning and Decontamination
Cleaning agents and tissues
To clean the GuardOne
®
Workstation, use the following cleaning agents and cleaning equipment:
lint-free cleaning tissue
alcohol-based disinfectant
For cleaning, a wipe disinfection with alcohol based disinfectants is recommended.
Further information can be found on the STARLAB homepage.
Cleaning interval
The following intervals must be adhered to:
After each use of the GuardOne
cleaning agent. Depending on the intended use of the GuardOne
by the UV lamp.
In case of a visible contamination, carry out appropriate cleaning.
To ensure the optimal function of the pre-filter, it is recommended to visually inspect the pre-filter in regular intervals. Exchange it as soon
as it is visibly dirty, latest after a period of 6 months.
Cleaning the outside area of the GuardOne
Preconditions:
The recommended cleaning intervals have been observed.
The GuardOne
®
Workstation is switched off or disconnected from the power supply.
1.
Use a lint-free cleaning tissue to dust off the outer glass chamber of the GuardOne
2.
If necessary, gently wipe the outside of the glass chamber with a mild detergent.
Cleaning the touchscreen
Preconditions:
The recommended cleaning intervals have been observed.
The GuardOne
®
Workstation is switched off or disconnected from the power supply.
1.
Use a lint-free, slightly damp cleaning tissue to gently wipe the touchscreen.
2
In case of heavy soiling, apply a little alcohol-based detergent to the cleaning tissue.
Gently wipe the touchscreen with the cleaning tissue.
42
· GuardOne
· Manual
®
®
Workstation, carry out the cleaning according to use. Clean the workstation with a recommended
Workstation
®
Activating the automatic life time surveillance
1.
To navigate to the Home page, press Tap to unlock.
2.
Press Settings.
3.
Press Replace UV Light.
4.
Enter the 10-digit serial number shown on the sticker on the packa-
ging of the UV lamp.
5.
Press Enter to finish the installation procedure.
The GuardOne
Workstation is now ready to be used again.
®
The remaining life time of the UV bulb is 9000 hours.
®
Workstation, carry out the decontamination of the PC glass chamber
Workstation.
®