8.1
Alerts
In case of an incoming alert, you can contact your local STARLAB sales representative or STARLAB service technician for further information
and assistance at any time.
Status alerts reflect information about the functionality and life time of the
Hepa filter and the UV lamp.
UV light life time end (Alert 1)
›
- UV light life time end means that the UV lamp has reached the end
time of 9000 hours and must be replaced to ensure the functionality
of the decontamination process.
›
UV light critical age (Alert 2)
- UV light critical age means that the UV lamp is close to the end of its
life time of 9000 hours but has not reached it yet. 10% of the life time
remains.
- The UV decontamination process can still be used but it is recommen-
ded to have a spare UV lamp ready for future replacement.
- To order a new UV lamp, contact your local STARLAB sales represen-
tative. Contact details may be found by pressing STARLAB Service &
Order Information on the Status page.
›
Hepa filter life time end (Alert 3)
- Hepa filter life time end means that the Hepa filter has reached its final
life time of 600 days and must be replaced to ensure the functionality
of the GuardOne
Workstation.
®
Hepa filter blocked (Alert 4)
›
- Hepa filter blocked means that the GuardOne
registered an impaired airflow that is usually caused by a blocked Hepa
filter.
- The Hepa filter must be replaced, irrespective of any remaining filter
life time.
- It is necessary to replace the Hepa filter immediately. Continuing the
operation is not recommended.
- Contact your STARLAB service technician for further assistance.
›
Hepa filter critical age (Alert 5)
- Hepa filter critical age means that the Hepa filter is close to the end
of its life time of 600 days but has not reached it yet. 10 % of the life
time remains.
- The functionality of the GuardOne
is recommended to have a spare Hepa filter ready for future replace-
ment.
- To order a new Hepa filter, contact your local STARLAB sales represen-
tative. Contact details may be found by pressing STARLAB Service &
Order Information on the Status page.
®
Workstation has
®
Workstation is not impaired but it
Fig. 54: Alerts
GuardOne
Troubleshooting
Alerts
45
®
· Manual ·