Environmental Guard Cleaning
5.6

Calibration Cap Cleaning Procedure

In general, the calibration cap should not require any maintenance. However, if it is
used in extremely dirty or solvent based environments, it may be necessary to clean
the light sensor if the calibration cap performance is not adequate.
Use a clean, dry applicator and isopropyl alcohol to gently swab the light sen-
sor, as shown in figure 21.
CAUTION
Never immerse the calibration cap in water or damage will occur.
Fig. 21
Calibration cap diode cleaning
US
PrimaX IR
Maintenance
37