Replacing The Fuse - Radiometer Analytical MeterLab PHM220 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

1.
Disconnect the PHM220. The 0.1 A fuse is positioned just above
the power socket on the rear panel.
2.
When the two vertical flaps on the cover are gently squeezed
together, the cover can be pulled out and the old fuse changed.
WARNING: For continued protection, only replace the fuse with
one of the same type and rating i.e. type T100 mA L250 V. For
compliance with UL/CSA standards, fuses recognised by these
bodies must be used.
3.
Re-insert the cover with the new fuse (450-010).
Note: Two secondary fuses are mounted on the printed circuit
board. If necessary, contact a Radiometer Analytical Service
Representative for replacement of the F1A, picofuse, UL
recognised.
58
Chapter 6

Replacing the Fuse