8. General information

Cleaning

The Titration Manager requires minimum maintenance. The exterior
surface can be cleaned with tepid water and wiped dry with a soft cloth.
Never use another solvent unless you have first consulted your Hach
representative.

Fuse replacement

For continued protection replace the fuse with one of a high interrupting
capacity, same type and rating:
2 x fuses, slow blow, 1.0 A (5 x 20 mm), part no. 450-020
To avoid the risk of electric shock, disconnect the device from the
power source before starting the fuse replacement process.
To replace the fuses:
1. Switch off the instrument.
2. Disconnect line cord.
3. Remove the fuse holder.
4. Replace the used fuses with ones of the same type and rating.
5. Put the cap back in place.
General information
77