Radiometer Analytical MeterLab PHM220 Mode D'emploi page 13

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Calibration Display Overview
AUTO
Buffers in use
IUPAC
Buffers in Use
4-7-10
Buffers in Use
TECHNICAL
Cal Temperature
25
˚C
Dip Electrode in
Buffer 1
STAB
4.57pH
CAL1
38s
25 ˚C
Buffer 1:
4.005
Dip in Buffer 2
STAB
6.99pH
CAL2 20s
25 ˚C
Buffer 2:
7.000
Dip in Buffer 3
STAB
10.01pH
CAL3 22s
25 ˚C
Buffer 2:10.012
Press
CALCULATING
PLEASE WAIT
ZeropH:
6.48
Sens.
: 99.3%
Auto printout if a printer is connected
Mode
Mode
Mode
pH
22MAR95 11:15:06
Cal
Cal
The buffer type recognised is displayed
for a few seconds
Appears when no temperature sensor
is connected. Adjust the temperature
manually using
Mode
Mode
ZeropH:
6.38
(Sens. : 100.3%)
Auto printout of a one-point
calibration if a printer is connected
FIXED
Cal Temperature
25
˚C
Dip Electrode in
Buffer 4.005
STAB
3.98pH
CAL1 38s
24 ˚C
Mode
Mode
Dip Electrode in
Buffer 7.000
STAB
6.99pH
CAL2 20s
25 ˚C
Dip Electrode in
Buffer 10.012
STAB
10.01pH
CAL3 28s
25 ˚C
CALCULATING
PLEASE WAIT
ZeropH:
6.48
Sens.
: 99.3%
Auto printout if a printer is connected
Starting up
13