In METHOD SETUP in the Edit mode, two printout format options are
available: ALL or RESULT.
1.
If ALL is selected: The instrument ID, the current date and time
are shown in the top line of the printout. The date when the last
calibration was performed, the zero pH and sensitivity results are
shown, followed by the sample number, the pH result and the
sample temperature. Finally, the stabilisation time is given:
*** PHM220 pH Meter 657R00N003
Last Calibration: 16FEB95
2 Point Calibration at 20.0 ˚ C: pH 7.000 (IUPAC) - 10.012 (IUPAC)
Zero pH:
7.02
Sensitivity:
Sample No.: 2
9.08 pH
Stabilisation time:
**************************************************************************
Note: If the Accept time has been set as the criterion in the Edit
mode (refer to Edit SAMPLE page 37), the Accept time
message will be printed instead of Stabilisation time . However,
if both an accept time and a stability criterion have been entered
and the result has been accepted manually by pressing , the
Stabilisation time will be printed out.
2.
If RESULT is selected: The instrument ID, the current date and
time are shown in the top line of the printout. The sample
number, pH result, sample temperature and stabilisation time are
then shown. As stated above, if Accept time has been set as
the criterion in the Edit mode; the Accept time will be printed
instead of Stabilisation time . An example of a RESULT
printout is shown below:
*** PHM220 pH Meter 657R00N003
Sample No.: 2
7.00 pH
Stabilisation time: 6s
**************************************************************************
Sample Results
10:15
100.8%
20.0 ˚ C
10s
25.0˚C
16FEB95
10:36 ***
16FEB95
15:25 ***
45
Printouts